4-17
VIBXPER
T II 05.2012
Configuring the result display (Display Setup)
In the Results display, after a measurement:
• Press the MENU key and click on 'Display Setup'.
In the selection field of the measurement tasks (Multimode):
• Highlight the measurement task symbol.
• Press the MENU key and highlight the 'Display Setup' entry.
• Press the navigation key right and click on 'Measurement' or 'Trend'
respectively (s. below).
X. Continuous measurement (‘live mode’)
In ‘Live mode’, you can check the signal quality before beginning with
data collection (see p. 3-4).
A1. Display Setup for measurement: Overall values
d
isplay
limit
as
...:
The limit values can be displayed as an absolute
value or as a difference ('Delta value') to the measurement value.
d
isplay
overall
peak
value
:
Select the characteristic vibration value that
is to be displayed by default with the 'RMS value'. After the mea-
surement, press the navigation key up repeatedly to display the
overall values of the series that were not displayed. If the thresh-
old value is exceeded, the RMS values and the overall value that
exceeded the threshold value by the largest amount are displayed.
A2. Display-Setup for Trend: Overall values
m
eas
.
value
:
Selection of
overall values
. The
line type
can be selected
and the
labeling
can be shown for each trend curve.
s
how
...:
Limits, event markers, references
and the associated
labeling
can be displayed in the chart. The
line type
can be selected.
Z
oom
: Speed or step size when zooming
l
ayout
: The values on the trend curve can be identified with
markers
.
A
grid
can be displayed for easier evaluation of the data and the
cursor can be displayed as a cross (
short
) or line cursor (
long
).
Left:
Opening the Display Setup
before a multimode measurement
Right:
Display Setup for Overall Trend
Display setup
Display Setup
for overall values with more than two
quantities.