3-24
VIBXPERT 11.2010
Options
Note
N
UMBER
OF
LINES
(S
PECTRUM
, C
EPSTRUM
): Number of lines displayed.
Together with the 'upper frequency', this parameter specifies the
resolution of the spectrum (
Δ
f).
W
INDOW
(S
PECTRUM
, C
EPSTRUM
): In the measurement of periodic signals,
there are time data sets with periodic gaps as a result of the finite
observation period. Thus, the frequency lines appear wider in the
corresponding spectrum. The window functions suppress these
'side lobes'.
Window features:
Hanning
: Standard for FFT. Frequency precise, but amplitude
errors (<15%); for the analysis of continuous processes with
hi
gh frequency resolution and lowest possible break-off effect.
Rectangular
: Frequency errors, but amplitude is precise; suitable
for the analysis of individual pulses, particularly if the pulse is at
the beginning of the time window.
Kaiser
: As with Hanning, but lower amplitude errors (<12%)
Flat-Top
: Lower frequency accuracy than Hanning, amplitude more
precise. For precision amplitude analysis of continuous processes.
Hamming
: Frequency errors as in Rectangular; amplitude errors
(18%) less than in Rectangular, but higher than in Hanning.
Bartlett
: Next to Rectangular, this is the simplest window (trian-
gle); lower amplitude and frequency accuracy.
Blackman
: As for Hanning, but with higher frequency accuracy;
amplitude errors as in Kaiser (12%); next to Kaiser, the second
best alternative for nearly all applications.
A
VERAGING
TYPE
/
NO
.
OF
MEASUREMENTS
(
SPECTRUM
, T
IMEWAVEFORM
, C
EPSTRUM
)
Without av.:
During the measurement each recorded spectrum is
displayed. The 'Number of averages" specifies how many single
measurements are carried out.
Linear:
The individual spectra are added and divided by the
number of averages. This averaging type emphasizes recurring
stationary sections in the spectrum.
Peak-hold:
Highest measurement value is displayed. This type of
averaging emphasizes peaks that occur stochastically.
Exponential:
Individual measurements are exponentially averaged
(the last spectrum is given the highest weighting).
Time-synchronous:
The individual measurements are averaged on
an RPM synchronous basis. Reference transducer required.
O
VERLAP
(
SPECTRUM
, C
EPSTRUM
): See section A1, P. 3-21.
M
EASUREMENT
RANGE
: See section A1, P. 3-22.
T
RIGGER
TYPE
/ L
EVEL
/ S
TART
(S
PECTRUM
, C
EPSTRUM
, T
IMEWAVEFORM
)
See section A1, P. 3-22
B. Setup parameters for sensors
The transducer parameters to be set up are described on P. 2-20. If
you have defined a default transducer for vibration measurements
(see p. 2-22), the ‘Sensor Setup’ menu is deactivated.