33
10
Maintenance
10.1
General
10.2
Maintenance instructions for the equipment for the operator
Despite the excellent quality of the units, you should carry out the following checks from time to time (1
- 2 months) for safety reasons:
•
Check all parts of the unit, especially parts subject to wear such as rollers, screws, bolts and
in particular the hydraulic resistance system, as well as their fastenings for strength and
damage. In order to maintain the technical safety level of the unit, parts damaged by wear
must be replaced immediately with original parts and the unit must be taken out of
operation until it is repaired.
•
If present, the locking pins and locking screws must also be greased.
With the hydraulic resistance transmitter, the following maintenance work is also required every 3-4
weeks:
•
Check all screw connections for tightness.
•
Check the resistance transmitters for leakage.
•
Clean the piston rods of the resistance transmitters and spray them with special oil.
•
Spray all spherical bearings (also those of the resistance transmitters) with special oil.
Notice!
If, contrary to expectations, you notice any damage, stop the training immediately and inform the
exercise instructor.
In case of device problems which you cannot solve yourself in a simple way, please inform the
proxomed
®
service in any case. The authorized service will help you quickly and competently or provide
you with instructions. Until the problem has been solved, the unit must be shut down and the shutdown
must be marked accordingly on the unit by the operator.
You can obtain the maintenance products at www.proxomedshop.de.
The training equipment of the compass®210 including the hydraulic
resistance system is subject to safety checks by the manufacturer or an
authorized service partner at least every 2 years.