The range of movement must be slow and controlled, while ensuring the correct execution of the movement. The
load intensity at the start of training should be deliberately selected to be low. The load is then increase gradually,
to avoid excessive loads, which can lead to problems and pain with patients/trainees.
Different training methods can be applied depending on the training and rehabilitation phase. The training intensity
is therefore based on the carrying capacity of the injured structure, the rehabilitation phase of the patient, and the
training goals.
Here priority can be given to the adaptation and habituation of the injured structure (connective tissue, cartilage,
tendons, ligaments, and joint capsule) to training with a load with low training intensity, before the training of
special power capabilities - such as stamina training or maximum power - with increasing intensity.
If discomfort occurs while training the exercise is to be stopped and a doctor is to be
consulted where necessary, as incorrect or excessive training can damage the health.
11.3
tergumed
®
710 Software Module
The tergumed
®
710
system can be combined with a software for measurement and training support. The software
module is available for all five tergumed
®
710 stations as an easy-to-use touch screen version. In addition, the
system includes a device-independent therapist workplace for the test evaluation and training planning.
The system supports the therapist/sports instructors by using isometric and dynamic tests in determining the deficit
in terms of the maximum force capacity and the development of training programs.
The support by the proxotrain tergumed
®
software module
does not, however, release the therapists from the
responsibility of selecting the right intensity for the patient or client individually and to take care of the patients
during the performance of the exercise. The detailed description of the software for the tergumed
®
710 line is
available in the corresponding User Manual for the software.
12
General settings and optional accessories
12.1
Fine adjustment 4 x 1 kg
Correct Usage
With the fine adjustment additional loads can be selected with levels of
0 kg, 1 kg, 2 kg, 3 kg, and 4 kg.
The fine adjustment function
The fine adjustment may only be adjusted in the unloaded condition (the
weight stack has no load).
By turning the knob of the fine adjustment, this can be set to the relevant
position and then it engages. The following positions are possible.
Position
4 x 1 kg
0
0 kg
1
1 kg
2
2 kg
3
3 kg
4
4 kg
According to the selected option, the set weight increase by the value of the fine adjustment.
12.2
Weight block
Correct Usage
The weight block on the tergumed
®
710 machines is always on the side or
attached in front of the trainee, this is so that the trainee or patient can select
the weight directly from the training position. The weights should always be
selected when not loaded. Additional weights are not to be attached, if this is
not authorised by proxomed
®
.
It is necessary that the weight pin be always inserted to until it stops. Only
original weight pins from proxomed
®
are to be used!
The weight block includes a special slot below the weight stack for fixing the
lever arm. By inserting the weight pin in this slot, it is possible to implement
isometric test and training procedures. An isometric test may take place only
under supervision.