Propaq CS Directions for Use
91
Maintenance
Inspect and Clean the Monitor and Accessories
Before cleaning, thoroughly inspect the monitor and all accessories for any signs of damage, cracks, or
improper mechanical function of keypads, switches, connectors, and printer paper door. While gently
bending and flexing cables and tubing, inspect for damage, cracks, cuts, abrasions, extreme wear,
exposed wires or bent connectors. Confirm connectors securely engage. Report damage or improper
function to your service department.
Caution
Do not autoclave the Propaq CS monitor or its accessories. Do not immerse the monitor in liquid when
cleaning. Do not immerse accessories in liquid when cleaning unless the accessory manufacturer’s cleaning
instructions explicitly instruct you to do so.
Equipment
Cleaning Instructions
Approved Cleaning Solutions
1
1. Do not use these cleaning solutions (they may damage the monitor): Butyl alcohol, Denatured ethanol, Freon™,
Mild chlorine bleach solution, Isopropyl alcohol, Trichloroethane, Trichloroethylene, Acetone, Vesphene II, Enviroquat
®
,
Staphene
®
, Misty
®
, Glutaraldehyde.
Propaq CS Monitor
2
2. The monitor may be disinfected to comply with OSHA requirements for cleaning and decontaminating spills of blood and
other body fluids. (Federal OSHA Standard on bloodborne pathogens: 29 CFR 1910.1030, 12/6/91.)
• Wipe with a nearly-dry cloth
moistened with cleaning solution.
• Thoroughly wipe off any excess
cleaning solution. Do not let water or
cleaning solution run into connector
openings or crevices.
3
3. If liquid gets into the right side panel connectors, it will drain out. If moisture gets into a left side panel connector, dry the
connector with warm air, then check the monitoring functions for proper operation.
Warm water Coverage
®
Liquid soap
Fantastik
®
Wex-cide
®
4
Formula 409
®
T.B.Q.
®
4
Windex
®
Cidex
®
Hydrogen peroxide solution
Ovation
®
4. Wex-cide (Wexford Labs, Inc., Kirkwood, MO) and T.B.Q. (Calgon Vestal Lab., Calgon Corp., St. Louis, MO) are
disinfectants that meet OSHA requirements, are EPA approved, and will not harm the outside of the monitor. Wipe away
disinfectants with a water-dampened cloth after the manufacturer’s recommended period of time.
NIBP cuff
• Wipe gently with cloth dampened
with cleaning solution.
• Thoroughly wipe off excess cleaning
solution. To avoid harming cuff
function, do not let water or cleaning
solution enter cuff tubing.
Common hospital disinfectants, including
Cidex, Clorox
®
liquid bleach (1:10 solution of
Clorox/water), isopropyl alcohol, Lysol
®
solution, Phisohex
®
, Quatricide
®
, Virex
®
and Vesphene
®
Cables, tubing, CO
2
sensor
5
5. The Mainstream CO
2
sensor may also be disinfected with Wex-cide. Follow the disinfectant manufacturer’s instructions. Do
not leave Wex-cide on sensor longer than 30 minutes. Thoroughly clean off residue with water-dampened cloth. Prolonged
exposure of the sensor to Wex-cide will damage the sensor.
• Wipe gently with cloth dampened
with cleaning solution. Do not immerse
the CO
2
sensor in liquid.
Mild detergent solution; also consult
manufacturer’s instructions.
Nellcor cables, Durasensor
oxygen transducers
• Wipe gently with cloth dampened
with isopropyl alcohol.
Isopropyl alcohol
Other accessories
• Consult manufacturer’s instructions.
Consult manufacturer’s instructions.
Содержание Propaq 242
Страница 6: ...6 Protocol Systems Inc...
Страница 18: ...18 Protocol Systems Inc...
Страница 26: ...26 Protocol Systems Inc...
Страница 50: ...50 Protocol Systems Inc...
Страница 52: ...52 Protocol Systems Inc...
Страница 64: ...64 Protocol Systems Inc...
Страница 66: ...66 Protocol Systems Inc...
Страница 72: ...72 Protocol Systems Inc...
Страница 74: ...74 Protocol Systems Inc...
Страница 78: ...78 Protocol Systems Inc...
Страница 84: ...84 Protocol Systems Inc...
Страница 86: ...86 Protocol Systems Inc...
Страница 94: ...94 Protocol Systems Inc...
Страница 120: ...120 Protocol Systems Inc...