Propaq CS Directions for Use
61
Alarms &
Aler
ts
Sidestream CO
2
Messages
ALTIMETER FAILURE - RANGE
. The Propaq CS monitor is operating at an altitude outside the
Sidestream CO
2
option’s operating altitude range of -2,000 to 15,000 feet. Returning the monitor to
within this range automatically cancels this message and restores operation.
ALTIMETER FAILURE - RATE
. The altimeter has detected that the ambient pressure is changing
at a rate greater than 100 mmHg/minute. When the rate of change is back within the 100
mmHg/minute range, disconnect and reconnect the CO
2
sensor to the monitor.
ALTIMETER NOT CALIBRATED - EEPROM -
The Sidestream CO
2
option has not been
calibrated. Refer the Propaq CS monitor to a biomedical engineer for calibration.
AMBIENT TEMPERATURE TOO HIGH
. The sensor temperature is too high. The Sidestream
CO
2
option is disabed until the ambient temperature is within the operating range specifications.
AMBIENT TEMPERATURE TOO LOW
. The sensor temperature is too low. The Sidestream CO
2
option is disabed until the ambient temperature is within the operating range specifications.
CALIBRATION ERROR - SERVICE REQUIRED.
Send the Propaq CS monitor to a biomedical
engineer for service.
DEGRADED WAVEFORM - SERVICE REQUIRED
. Send the Propaq CS monitor to a biomedical
engineer for service.
LACK OF WAVEFORM - SERVICE REQUIRED
. Send the Propaq CS monitor to a biomedical
engineer for service.
MOTOR FAILURE - SERVICE REQUIRED.
The sensor hardware has failed. Send the Propaq CS
monitor to a biomedical engineer for service.
NO WATERTRAP DETECTED
. There is no Sidestream CO
2
watertrap installed. Install a
watertrap.
OCCLUSION - CHECK EXHAUST PORT/TUBING
. Blockage has been detected on the
pneumatic exhaust port. Check the exhaust port and related tubing for occlusions. Make sure that
the sampling line and any inputs to the patient breathing apparatus are not connected to the
exhaust port.
OCCLUSION - CHECK WATERTRAP/TUBING
. Blockage has been detected on the Sidestream
CO
2
input. Check the watertrap, sample line, and any connected tubing for occlusion.
PUMP FAILURE,
SERVICE REQUIRED. The pump is not able to maintain the target flow rate.
Send the Propaq CS monitor to a biomedical engineer for service.
SIDESTREAM STICK EEPROM FAILURE
. Send the Propaq CS monitor to a biomedical
engineer for service.
SSP BOARD EEPROM FAILURE
. Send the Propaq CS monitor to a biomedical engineer for
service.
The following messages can appear in the numerics display area.
OFF.
No CO
2
source is selected.
SRCH.
The sensor is preparing for a measurement.
START UP.
Sidestream CO
2
has been activated and is preparing for operation. This typically
requires 30 seconds at room temperature.
UNCAL.
The monitor has detected a problem such as a lack of calibration, an obstruction, or a low
battery.
Содержание Propaq 242
Страница 6: ...6 Protocol Systems Inc...
Страница 18: ...18 Protocol Systems Inc...
Страница 26: ...26 Protocol Systems Inc...
Страница 50: ...50 Protocol Systems Inc...
Страница 52: ...52 Protocol Systems Inc...
Страница 64: ...64 Protocol Systems Inc...
Страница 66: ...66 Protocol Systems Inc...
Страница 72: ...72 Protocol Systems Inc...
Страница 74: ...74 Protocol Systems Inc...
Страница 78: ...78 Protocol Systems Inc...
Страница 84: ...84 Protocol Systems Inc...
Страница 86: ...86 Protocol Systems Inc...
Страница 94: ...94 Protocol Systems Inc...
Страница 120: ...120 Protocol Systems Inc...