PRS 500 C 

 

Basic diagnostic X-ray system 

 

Model/ID: 7073-9-8050 

 
 

 

Instructions for use 

 

Ident. No. 5073-0-3002 

 

 

Содержание PRS 500 C

Страница 1: ...71720 Oberstenfeld Germany Telephone 49 0 7062 92 55 0 E Mail protec protec med com Version 4 0 Issued 2021 03 11 Subject to alterations PRS 500 C Basic diagnostic X ray system Model ID 7073 9 8050 In...

Страница 2: ...uthorization of PROTEC s law department Report any violations of this requirement to PROTEC GmbH Co KG 2020 PROTEC GmbH Co KG Oberstenfeld These accompanying documents were created and distributed by...

Страница 3: ...ation protection 16 2 1 7 Ventilation 16 2 1 8 Interaction with external devices 16 2 1 9 Electromagnetic Environment and the influence of devices 17 3 Control elements and device displays 18 3 1 Cont...

Страница 4: ...the unit operation by the user 26 5 3 2 Regular controls by the user 26 5 3 3 Only original spare parts are to be used in situations requiring component replacement Maintenance 26 5 3 4 Warranty 26 5...

Страница 5: ...PRS 500 C Instructions for use 5073 0 3002 5 von 35 8 3 Labels 34 8 4 Position symbols and labels 35 8 5 Abbreviations 35...

Страница 6: ...ages Comments 1 0 2020 11 12 all Original issue 2 0 2021 02 17 7 23 Radiation Warning Power Box 3 0 2021 02 25 Front Page 10 14 18 21 24 30 34 Product picture Warning mattress mattress at characterist...

Страница 7: ...tics X rays are ionizing radiation that can cause damage to living organisms e g cancer or mutations X rays represent a potential risk for patients and employees Therefore the aim is to minimize radia...

Страница 8: ...sidual risk will remain in clinical use These are displayed in the following user manual by application limitations contraindications warnings and precautions The installation and service of equipment...

Страница 9: ...ographic table fixed biocompatible X ray mattress This mattress is specially designed for X ray imaging diagnostics The mattress can be removed for cleaning Floor railed X ray system tube support floo...

Страница 10: ...th anti scatter grid and 3 field measuring chamber intended for the use with automatic exposure control Useable for variable cassette detector sizes Formats from 13 cm x 18 cm 5 x 7 to 43 cm x 43 cm 1...

Страница 11: ...o answer the diagnostic question and no other more suitable alternative method is available Accordingly it is also described by the International Atomic Energy Agency IAEA in the document Radiation Pr...

Страница 12: ...ional who are trained in accordance with the corresponding national regulations in the use of diagnostic X Ry equipment and its proper certified use in connection with other medical products objects a...

Страница 13: ...ian NOTE Every delivered manual has to be read and the safety notes have to be observed NOTE After installation the commissioning have to be recorded with the PROTEC acceptance protocol NOTE For the d...

Страница 14: ...lfilled The X ray system has to be checked according to the local law and also accepted by the responsible office CAUTION If the wrong SID is in use for exposures personal injuries for the patient can...

Страница 15: ...e PRS 500 C from the power the connected X Ray generator must be shut off WARNING The positioning and support of a patient on the X ray system table must not be carried out without an X ray mattress 2...

Страница 16: ...ouble the distance dose triple the distance 1 9 dose Keep the exposure time as short as possible The dosage is directly correlated with the exposure time A half exposure time results in a radiation do...

Страница 17: ...avoided since it can cause an improper operating mode If there is no other possibility than this the PRS 500 C and other devices should be studied to make sure they work proper NOTE The characteristic...

Страница 18: ...position horizontally it can be moved manually while holding this button UP DOWN When operator wants to move the tube arm vertically at desired position it can be moved manually while holding this but...

Страница 19: ...oot or any object in this area the ALL FREE switch will be activated With the switch checked it is available to move the table top in any direction 3 1 3 Vertical Bucky Wall Stand UP DOWN lock Release...

Страница 20: ...scales Pos 7 Format width scales Pos 8 Tape measure for SID measurement Read off the measurement on the bottom edge of the collimator optional Pos 9 Two accessory rails Pos 10 Locking spring to fix ac...

Страница 21: ...engaging the cassette positioner in the size of the cassette corresponding detent 13 cm 18 cm 24 cm 30 cm 35 cm 40 cm or 43 cm push the cartridge to the cassette positioner Push the cassette clamps ag...

Страница 22: ...ky Wall Stand 4 1 2 2 Adjustment of the source to image receptor distance SID Release the longitudinal movement brake of the column by pressing button Rail Slide and adjust the source to image recepto...

Страница 23: ...anual of the X ray tube 4 4 Operation X ray generator Detailed information please find in the enclosed User Manual of the generator 4 5 Operation Bucky Grid entity Detailed information please find in...

Страница 24: ...switch you could the power line of Generator and 230V components switched off and on see picture 4 8 2 Dosimetric Calibration VENUS CMP RFX generators Detailed information please find in the enclosed...

Страница 25: ...ress is equipped with a Velcro system To make cleaning easier the mattress can be removed to clean the areas underneath and the mattress itself After cleaning the X ray mattress must be reattached to...

Страница 26: ...tuations requiring component replacement Maintenance The required maintenance must be carried out by PROTEC technical service or an authorized service provider to ensure the safe and reliable function...

Страница 27: ...system and of the individual components 5 3 7 Applied Parts and parts which get handled like an application part Part Definition as applied part or parts which get handled like an application part bu...

Страница 28: ...ENUS 50R 1 phase VENUS 32R 3 phase VENUS 32R 1 phase Output Power 50kW 50kW 32kW 32kW Power supply voltage 380V AC 400V AC 220V AC 380V AC 400V AC 220V AC 230V AC Phase 3PH N PE 1phase 3PH N PE 1phase...

Страница 29: ...se a decrease in the performance features of the device CAUTION The X Ray generator integrated into the radiographic system PRS 500C sends out electromagnetic waves during operation which could cause...

Страница 30: ...PRS 500 C Instructions for use 5073 0 3002 30 von 35 7 Technical Data 7 1 Dimensions PATIENT PATIENT PATIENT PATIENT OP OP OP...

Страница 31: ...idth 800 mm Height 15 mm Patient load max 200 kg Velcro fastening cleanable biocompatible 7 3 Bucky unit Longitudinal travel 580 mm Table top film distance 75 mm The Bucky unit and the measure chamber...

Страница 32: ...st half value layer thickness HVLT of 2 7mm Al The X ray mattress is defined as an applied part The attenuation equivalent of the X ray mattress is typically 0 5 mm Al to EN 6061 1 3 at 100 kV with a...

Страница 33: ...ed weight Do not exceed the maximum indicated weight Table top movements for exposure Longitudinal movement of the table top Transverse movement of the table top The positioning and support of a patie...

Страница 34: ...nd fingers while moving the table top table and or X Ray tube assembly unit Maximum safe working load radiographic table without X ray mattress Labels on the X ray system tube support floor stand and...

Страница 35: ...y mattress The label is located on the underside of the X ray mattress 8 5 Abbreviations mm Millimetre cm Centimetre lb Pound kg Kilogram C Degree Celsius hPa Hectopascal DIN German Industry Standard...