User Manual
PAR-809C
VARIABLE RESISTANCE REFERENCE
Страница 1: ...User Manual PAR 809C VARIABLE RESISTANCE REFERENCE ...
Страница 2: ......
Страница 3: ...nance 9 General Specifications 10 Table of Contents Copyright 2011 by Prostat Corporation All rights reserved Printed in the United States of America No part of this manual may be used or reproduced in any manner whatsoever without written permission For information contact Prostat Corporation 1072 Tower Lane Bensenville IL 60106 USA Prostat is the registered trademark of Prostat Corporation ...
Страница 4: ...lues should be conducted annually II Cautions Warnings A Do not use the PAR 809C on circuits exceeding 100 volts DC B As with any electrical device use proper electrical precautions to avoid personnel shock C Only qualified instrument repair personnel should open or repair the PAR 809C D Do not store or use in damp environments E PAR 809C Equipment and Accessories 1 One PAR 809C Variable Resistanc...
Страница 5: ...e confirmation of critical ESD control tools V General Operations A Confirming Wrist Strap Tester Resistance Performance Range 1 Plug the two supplied test leads into the PAR 809C receptacles 2 Plug one PAR 809C test lead into the wrist strap tester s Wrist strap Ground Cord receptacle 3 Rotate the PAR 809C s Position Selector knob to Position 1 the lowest resistance reference 4 Turn ON the PAR 80...
Страница 6: ...ce under test as too low b Resistance positions above the PASS range should all be rejected by the device under test as too high B Confirming Foot Wear Tester Resistance Performance Range 1 Plug the two supplied test leads into the PAR 809C receptacles 2 Attach an alligator clip to one PAR 809C test lead Attach the alligator clip to the Foot Wear test plate 6 Rev B January 2012 PAR 809C Variable R...
Страница 7: ...e Foot Wear device under test 7 Vary the PAR 809C reference resistance by rotating the Position Selector Knob on the face of the PAR 809C to the next position and repeat step 6 8 The resistance range positions where the device under test indicates PASS are those resistance values accepted by the Foot Wear tester a Resistance positions below the PASS range should all be rejected by the device under...
Страница 8: ...tion tested D Confirming Functional Performance of the PWA 805 Wrist Strap Auditor NOTE The PRS 812 or PRS 801 is used for this functional test procedure 1 Check the BATTERY function of the Resistance Meter replace batteries if indication is LOW 2 Turn the Resistance Meter to its OFF position 3 Install the meter test leads into the proper receptacles 4 Plug the tow test lead banana plugs into the ...
Страница 9: ...yed in this fixture VI Maintenance A Replace 9 volt battery at least once each year Remove battery when device is in extended storage i e six months or greater B Periodically confirm resistance values of each resistance position at 10 volts using precision wide range ohmmeter Contact Prostat Corporation for calibration Service if necessary C Resistance measurement confirmation should be conducted ...
Страница 10: ...n Rotary Range selector knob selects reference resistance Knob Height 1 0 ON OFF Slide Switch controls 12 LED position indicator lights Connections 2 Banana receptacles at end of case Used to connect device under test in series with selected resistance Indication 12 LED Range Position indicator lights Accessories 2 each 24 min output cables with standard male banana plugs at each end Power One 9 V...
Страница 11: ......
Страница 12: ...Prostat Corporation All other trademarks and trade names are the property of their respective companies P R O F E S S I O N A L S T A T I C C O N T R O L P R O D U C T S Prostat Corporation Corporate Headquarters 1072 Tower Lane Bensenville IL 60106 630 238 8883 Fax 630 238 9717 1 855 STATIC1 www prostatcorp com ...