7
Operation
7.1 Calibrating DULCO
®
turb C
The DULCO
®
turb C is tested and
calibrated before leaving the factory.
For this reason, the DULCO
®
turb C
may be used immediately. Under
normal conditions, it is recommended
to carry out calibration at least once
every three months.
If the DULCO
®
turb C is in [CAL] or
[CONFIG] mode, the alarm relay
changes to alarm status. If no input is
made in [CAL]mode within a period of
15 minutes, the DULCO
®
turb C auto‐
matically switches back to [AUTO]
mode.
7.1.1 Standard solutions for cali‐
bration
If the DULCO
®
turb C is operated
over its complete measuring range of
0.02 NTU ... 1000 NTU, you must per‐
form calibration with all 3 standard
solutions (0.02 NTU, 10.0 NTU and
1000 NTU). If the sensor is only oper‐
ated in measurement ranges under
10 NTU, then you may perform a sim‐
plified calibration with standard solu‐
tions 0.02 NTU and 10.0 NTU.
CAUTION!
Use of diluted formazine
Possible consequences: Inaccurate
calibration. Malfunctions in dependent
processes.
Diluted formazine is instable. Ensure
that freshly prepared formazine solu‐
tion is used when performing calibra‐
tion.
If possible, use a standard solution
from Prominent for calibration. These
standard solutions are more stable
than formazine and have a minimum
shelf-life of 12 months. In doing so,
observe the use-by-date on the
standard solution packaging.
Calibration set order number: 1037699
Operation
55