4.2 Installation (hydraulic)
CAUTION!
Free flow at outlet
Direct any sample water that has
flowed through the DULCO
®
turb C
into a free outlet. The sampled water
may not be fed back into the process.
CAUTION!
Algae growth under the influence of
light
Possible algae growth if unsuitable
hoses are used.
Do not use transparent hoses if the
installation is subjected to powerful
sources of light. This prevents the for‐
mation of algae in the installation.
Leakage on the vent screw
Slight leakage may occur at the bore
hole for the vent screw during the
installation process. This will cease as
soon as normal flow is established.
In the event that the installation con‐
tinues to leak permanently at this
point due to the high pressure in this
area, we recommend sealing off the
bore hole with the sealing plug sup‐
plied. In order to ensure free
drainage, a bleed option must be
installed at another area of the outlet
in order to avoid overpressure or
underpressure in the system.
Permissible operating parame‐
ters
–
Maximum liquid temperature 50
°C
–
Maximum pressure 13.8 bar
–
Flow rate 6 ... 60 l/h
Assembly and Installation