Ambient temperature: +5 ... +50°C
Medium temperature: max. sample water temperature 45 °C
Medium temperature: max. sample water temperature 65 °C, for
sample water “H”.
Medium temperature: max. sample water temperature 35 °C, for
the measured variable fluoride.
Air humidity: max. 90 % relative air humidity, non-condensing.
Degree of protection: IP 65
The transport weight depends on the configuration of the meas‐
uring plate and lies between 8 ... 13 kg.
Free flow / No overpressure or negative pressure
Make sure that there is no overpressure > 1 bar
and also no negative pressure at the outlet of the
measuring plate. Overpressure damages the
bypass fitting. Negative pressure damages the
sensors fitted. Use the stopcocks to adjust the
pressure at the sample water inlet and outlet.
Route the pipework so that the sample water can
drain unimpeded.
Temperature data
Climate
Protection against contact and mois‐
ture (IP)
Transport weight
Technical data and operating parameters