8. Tools

Use the Tools menu to access Settings and Diagnostics. Touch the 

Tools

 key to reach the menu.

8.A. Settings

Contrast

The screen contrast adjustment is located under the Settings menu. Use the left and right arrow keys to increase the 
brightness of the touch screen. Touch 

Home

 when finished to return to the Home Screen.

Reset

The reset key is located under the Settings menu. It automatically power cycles the luminometer. The power cycle 
erases any unsaved protocols and the data on the Home Screen. The reset key is an easy way to restart the luminometer 
after inserting a Fluorescence Module or the Light Standard.

Lid Start

The lid start feature is also located under the Settings menu. When lid start is enabled, the luminometer begins 
measurement immediately after closing the lid. Touch 

Lid Start

 once to enable the lid start feature. Return to the 

Tools menu to disable the Lid Start feature.

8.B. Diagnostics 

Touch Screen Calibration

Access the screen calibration procedure from the Diagnostics menu. Although the touch screen is calibrated at the 
factory, it may need recalibration during instrument setup and over time. Contact Promega for more information on 
screen calibration. If the touch screen appears nonresponsive and the Diagnostics menu cannot be activated, please see 
the alternative instructions below.

1. Select 

Touch Screen Calibration

.

2. 

When the screen reads “Please touch the calibration point”, touch the gray box on the + with a rounded object 
like a pen cap, a stylus or the pointed end of a microcentrifuge or PCR tube. Do not use a pen or anything sharp 
because this might damage the screen.

3. 

A second gray box will appear when the first + has been touched. Touch the 

+

 in this gray box.

4. 

Options for saving the calibration will appear. Touch 

Apply New Calibration

 if the new calibration is correct. 

Select 

Reset To Factory Calibration

 or 

Abort Calibration

 if a mistake is made during the calibration process.

 

Alternatively, the touch screen can be reset to factory calibration settings by powering off the instrument and 
then placing your finger on the screen during power on.

 

Keep your finger on the screen until the Home Screen appears. The touch screen should be reset to factory settings.

Device Configuration

The 

Device Configuration

 key contains information on the firmware version and other useful information for 

troubleshooting instrument-related problems.

Содержание E5311

Страница 1: ...Revised 5 21 TM276 T E C H N I C A L M A N U A L GloMax 20 20 Luminometer Instructions for Use of Products E5311 E5351 and E5361...

Страница 2: ...ct Components 4 3 Hardware Overview 4 4 Setting up the GloMax 20 20 Luminometer 5 4 A Setup 5 4 B Software Installation and System Connection 5 5 Operating the GloMax 20 20 Luminometer 6 5 A Touch Scr...

Страница 3: ...USA 800 356 9526 608 274 4330 Fax 608 277 2516 TM276 Revised 5 21 www promega com 10 Troubleshooting 14 11 Appendix 15 11 A General Care 15 11 B Warranty and Service 15 11 C Specifications 16 11 D Ce...

Страница 4: ...horseradish peroxidase and alkaline phosphatase The touch screen on the luminometer allows easy setup and operation The GloMax 20 20 Luminometer also features optional modules for measuring certain fl...

Страница 5: ...311 Includes 1 Luminometer 1 1 5ml Microcentrifuge Tube Holder 1 USB Cable 1 Power Supply Kit 1 Quick Start Guide FB081 Items Available Separately P R O D U C T S I Z E C AT GloMax 20 20 Fluorescent M...

Страница 6: ...GloMax Spreadsheet Interface Software is compatible with 32 and 64 bit PCs using Windows 7 Professional Service Pack 1 with either Microsoft Office 2010 Home and Business Edition 32 and 64 bit versio...

Страница 7: ...a function touch the button corresponding to the function once 5 B Home Screen The Home Screen Figure 3 provides orientation for the multiple functions of the GloMax 20 20 Luminometer From the Home S...

Страница 8: ...surement settings automatically programmed Touch OK to accept the parameters and return to the Home Screen You are now ready to measure your samples 5389TA Figure 4 The Protocol Menu 5 E Creating a Ne...

Страница 9: ...tocol To access a saved protocol select Run User Protocol The next screen will display protocols previously saved on the luminometer Identify and touch the name of your protocol The next screen will d...

Страница 10: ...roll through recent measurements Note Measurements are not saved after power down The instrument display will only show 20 measurements additional measurements will overwrite the first measurements Fo...

Страница 11: ...nu Although the touch screen is calibrated at the factory it may need recalibration during instrument setup and over time Contact Promega for more information on screen calibration If the touch screen...

Страница 12: ...e Identification Screen will appear Identify the Light Standard The luminometer will automatically power OFF The GloMax 20 20 Luminometer will automatically identify the Light Standard for subsequent...

Страница 13: ...ule is designed for easy installation and removal Store the Fluorescent Module in the product case when not in use Keep liquids dust and debris away from direct contact with the Fluorescent Module 9 A...

Страница 14: ...ppb ppt or FSU With a valid calibration it is not necessary to generate a standard curve to determine the actual concentration of the sample Calibration Touch Calibration to calibrate the Fluorescent...

Страница 15: ...reen until the Home Screen appears Also see Section 8 B Touch Screen Calibration Excel does not open Excel is not installed on the PC Make sure Microsoft Excel is installed on your PC and activated Th...

Страница 16: ...11 B Warranty and Service The GloMax 20 20 Luminometer comes with a one year warranty from Promega Additional warranty is available For more information contact Promega Technical Services Contact info...

Страница 17: ...ft Office 2007 32 bit version Power 0 5A 100 240V 50 60Hz universal Dimensions 12 92 D 10 44 W 8 42 H 32 82 26 52 21 39cm Weight 8 1 lb 3 65kg Warranty One year CE Compliance Pollution degree 2 Instal...

Страница 18: ...on will result in decontamination charges before the instrument will be serviced Select either A or B A I confirm that the returned items have not been contaminated by body fluids or by toxic carcinog...

Страница 19: ...20 Replacement Power Supply 1 each E5411 Thermal Serial Printer and Cable Universal Power 1 each E2821 Tripp Lite USB to Serial Adapter 1 each TS2520 Microplate Readers Product Size Cat GloMax Discove...

Страница 20: ...rporation Microsoft and Windows are registered trademarks of Microsoft Corporation PicoGreen and RiboGreen are registered trademarks of Molecular Probes Inc Products may be covered by pending or issue...

Отзывы: