19
Promega Maxwell
®
CSC 48 Instrument | TM623 | Revised 6/20
4.3 Shutting Down the Maxwell
®
CSC 48 Instrument
Shutting Down Instrument
1. Shut down the software by pressing the
X
in the upper left corner of the ‘Home’ screen.
From any other screen in the software, touch the
Home
button at the upper left corner of
the screen to return to the ‘Home’ screen.
2. Turn the Maxwell
®
CSC 48 instrument off by pressing and holding the power button on the
right front side of the instrument for 3 seconds. Switch the rocker switch on the back of the
instrument to the off position. Unplug the instrument. If you need to store the instrument,
after following the steps above, store in a place that meets the environmental requirements
described in Section 1.10.
3. Shut down the Tablet PC using the Windows
®
10 shut-down procedure: Touch the
Windows
®
icon in the lower left corner of the screen, touch the power icon on the left side of
the Windows
®
menu and choose Shut Down.
Storing the Tablet PC
When not in use for a prolonged period of time, the Tablet PC should be unplugged.
4.4 Configuring the Tablet PC
The Maxwell
®
CSC 48 Instrument is controlled by Maxwell
®
CSC 48 software running on a
Tablet PC. The Tablet PC should be configured to meet the needs of your site, including setting
date and time, adding users, specifying access levels for users, connecting to a network and
adding network printers. Instructions for configuring the Tablet PC can be found in the
Maxwell
®
CSC Tablet PC Configuration Manual
#TM484.
Your IT department
or site Administrator
should configure the
Tablet PC according
to the IT rules and IT
procedures pertinent
to your site.
Preparing the Maxwell® CSC 48 Instrument for Use
Содержание AS8000
Страница 66: ...65 Promega Maxwell CSC 48 Instrument TM623 Revised 6 20 Notes Operating the Maxwell CSC 48 Instrument ...
Страница 70: ...69 Promega Maxwell CSC 48 Instrument TM623 Revised 6 20 Notes Cleaning and Maintenance ...
Страница 74: ...73 Promega Maxwell CSC 48 Instrument TM623 Revised 6 20 Notes ...
Страница 86: ...85 Promega Maxwell CSC 48 Instrument TM623 Revised 6 20 Notes ...