October 2014
Page
13
CAUTION:
In order to avoid suffering from the electric shock, please keep your
finger behind the safety guard ring of the probe body during the
operation. In order to protect you from suffering from the electric shock
during your using the probe, do not touch the metal part of the probe tip
when the probe is connected to the power supply. Before making any
measurements, please connect the probe to the instrument and connect
the ground terminal to the earth.
2.2.7
How to Implement Auto-calibration
The auto-calibration application can make the oscilloscope reach the optimum
condition rapidly to obtain the most accurate measurement value. You can carry
out this application program at any time, but when the range of variation of the
ambient temperature is up to or over 5
ºC
, this program must be executed.
For the performing of the self-calibration, all probes or wires should be
disconnected with the input connector first. Then, press the "
8(UTILITY)
"
button to call out the
FUNCTION
menu; push down the F3 menu selection
button to choose the option " Do Self Cal"; finally, run the program after
confirming that everything is ready now.
2.2.8
Introduction to the Vertical System
Shown as Fig.10, there are a series of buttons and knobs in
VERTICAL
CONTROLS
. The following practices will gradually direct you to be familiar with
the using of the vertical setting.
Figure 10
. Vertical Control Zone.