Setting up
With a short press of the on/off button
the device is switched on or off.
You will see the Zehntner-logo and the firmware version during switching on before
the display changes into measuring mode.
For switching off the instrument you have to press the on/off button
.
6
Calibrate
6.1 Calibrate
To ensure precise and valid results the ZRM 6006 must be calibrated by the user
on the working standard in regular intervals (recommended interval: every 2 days).
Put the ZRM 6006 on the working standard, switch it on and press the measuring
button longer than 5 seconds until the calibration values are shown on the display.
Release the measuring button. The instrument is in calibration mode now. To
execute a calibration, press the measuring button shortly.
If the measuring button is not pressed within 2 seconds,
the instrument switches back to measuring mode and
the main screen will be displayed.
If the calibration is successful, a check
will be shown
behind the calibration value. If the calibration value differs from the factory
calibration by more than 15% an X will be displayed.
Values with high deviation will not be accepted
automatically. If the text “PRESS M TO ACCEPT ANY”
is displayed and the measuring button is pressed the
calibration out of range can be accepted.
If an error occurred during calibration, “-“ will be
displayed behind the calibration value. In this case it is
not possible to accept the calibration value. Please
contact Proceq SA or your authorized Proceq-agent.
For calibration the device must be positioned
exactly on the standard. Make sure that the snap-lock mechanism of the
standard is attached correctly to the instrument.