6.2.3 Mounting the friction finger set C
1) Check the cellular rubber for
damages.
2) Place the rubbing cloth on the cellular
rubber with one corner (circled in red)
facing the holder screw (circled in green).
3) Clamp the rubbing cloth smoothly. 4) Slide on the weight and screw it in place.
If your friction finger set does not look like picture 3, meaning one corner of
the rubbing cloth (circled in red) does not face the holder screw (circled in
green), you need to redo step 2.
The test load of 9 N will only be reached if the knurled nut is mounted.
To avoid ejection of weights (accidentally falling down) they have to be
screwed to the holder.