Pro-Ject Audio Systems Stylus Balance Скачать руководство пользователя страница 1

© Audio Trade GmbH · Measure it II · Revision 15.06.11/15.07.11

 

 

Thank you for purchasing a PRO-JECT AUDIO stylus balance. 

!

 

Warning of a hazard for the user, the unit or possible misuse.  

 

Important notice. 

Calibration 

Make sure the surface is level.

 

2

 Turn on the meter with the 

ON/OFF

 button. The display will show 

0.000

 after a few seconds. Give your 

stylus balance sufficient warm up time. Usually 30 seconds to give the internal components time to stabilise 
before calibrating. 

3

 Press and hold the 

MODE

 button until the display shows 

CAL

 and 

5.000 

(flashing). 

4

 Gently place the 5g weight on the centre of the platform. After approximately 5 seconds the display will 

read 

5.000

PASS

 and 

5.000

 

Occasionally the display will show for example 

5.002

 or 

5.004

 instead of 

5.000

. This is no defect. 

It will not affect the calibration.

 

5

 Remove the weight an press 

ON/OFF

 to turn off the meter.

 

Choosing the weighting mode 

The meter has to be turned on. The display shows 

0.000

. Pressing the 

 MODE

 switches between the different 

weighting modes.  
Weighting modes: gram (1g = 10mN), ounce (oz), pennyweight (dwt) and carat (ct). 

Cartridge downforce adjustment - weighting mode gram

 

Make sure the surface you use the turntable on is level.

 

2

 Remove the stylus guard and the stylus balance protection cover. Gently place the meter on the platter. 

Move the headshell over the weighting platform. 

3

 Turn on the stylus balance with the 

ON/OFF

 button. The display will show 

0.000

 after a few seconds. 

Give your meter sufficient warm up time. Usually 30 seconds to give the internal components time to 
stabilise. 

4

 Gently lower the tonearm until the stylus is on the black dot of the weighting platform. The display shows 

the tracking force. 

5

 Lift the tonearm and adjust the counter weight (tracking weight dial) and repeat the above procedure the 

recommended tracking force is achieved. 

!

 

Never load the stylus balance with more than 

5g

, overloading can permanently damage it.

 

Other display readouts 

No display, backlight getting weak - batteries are empty and need to be replaced. 

o_LD

 - overload (more than 5g). Please remove the excessive load immediately. 

LLLLL

 - The weighting load cell has been damaged by misuse (i.e. it was shocked, dropped or overloaded). 

You can try to recalibrate the meter. If the load cell has not been damaged to much, it will work again after 
recalibration. 

UNSt

 - Use the meter on a more stable, vibration free surface. 

Backlight 

The 

LIGHT

 key switches the backlight off an on. 