 GENERAL INSPECTION AND MAINTENANCE 

 

No maintenance is to be carried out on the chair while in use to reduce the risk of injury. 

 

 

The user will not require any tools when completing their inspections. 
When completing services, the authorised service engineer will provide their own tools, including: 

 

1 x set of metric Allen keys. 

7.1  Daily Checks - To Be Completed by User 

Prior to each use the Combi toileting chair must be inspected visually.  

 

Should any of these items fail the inspection, DO NOT use the chair. 

Contact your local authorized dealer for service – contact details are on the last page of 

this manual. 

 

Visually check for the following: 

 

  Check the tray (if supplied) to ensure no parts are loose or damaged. 

  Check all upholstery for signs of wear and tear. 

  Check that the hand wheels tighten correctly, and are not cross-threaded or damaged. 

  Check that the leg screws are present and are tightened correctly. 

  Check all straps for fraying, and that buckles are not missing/damaged. 

  Check the strap is securely fastened to the poppers on the underside of the chair. 

7.2

 

 Service Interval 

The Combi toileting chair should be serviced every 3 years. Servicing must only be undertaken by 

a Smirthwaite Ltd. service engineer, or by a Smirthwaite Ltd. trained representative.   

7.3

 

 Nominal Service Life 

Your product has a nominal service life of 12 years, during which full post-sales support will be 

available with regard to spares and servicing. 

 

Product service life has been determined based upon the design complexity of product, and the 

anticipated exposure to normal use.   

 

Good practice dictates all Smirthwaite Ltd. products have been designed and manufactured to high 

levels of safety and quality, and will meet requirements of normal use when maintained in line with 

our servicing recommendations. 

    

 

Содержание smirthwaite Combi

Страница 1: ...1 P a g e User Manual Combi...

Страница 2: ...Parts 8 2 2 Packaging 9 2 3 Chair Assembly When Supplied Flat Packed 9 3 0 Installation 10 3 1 Adjustments 10 3 1 1 Seat Depth 10 3 1 2 Seat Height 10 3 1 3 Armrest Height 11 3 1 4 Pelvic Strap 11 3 2...

Страница 3: ...nance 19 7 1 Daily Checks To Be Completed by User 19 7 2 Service Interval 19 7 3 Nominal Service Life 19 7 4 Extending Nominal Service Life 20 7 5 User Serviceable Parts 20 7 6 Final Inspection 20 8 0...

Страница 4: ...ents while reviewing this manual Contents of this manual are subject to change without prior notice Unauthorised modifications on any Smirthwaite product may affect its safety The manufacturer will no...

Страница 5: ...aws on recycling Degree of protection provided by enclosure N1 Ingress of particles N2 Ingress of water Temperature range Humidity range Type B applied part Type BF applied part Medical Device Authori...

Страница 6: ...eight and depth 1 6 Chair Operating Environment The Combi toileting chair is suitable for use within the professional health care facility environment as well as the home health care environment The c...

Страница 7: ...raps provided with the chair Regularly check that straps and or harnesses are not showing signs of tearing or fraying and that they fasten correctly DO NOT use with bath oils or bath salts If any part...

Страница 8: ...ire the chair to be fully assembled which takes approximately 2 minutes If in any doubt always seek advice Always turn hand wheels levers and screws clockwise to tighten or anti clockwise to loosen Co...

Страница 9: ...rrectly do not use until rectified 2 3 Chair Assembly When Supplied Flat Packed Using the screws and tool provided attach the seat plate to a seat side by inserting two screws Do not tighten the screw...

Страница 10: ...osening the two screws G located on each side of the chair Adjustment range is limited by the slot length on each side of the chair Ensure each screw is tight after adjustment and before placing the c...

Страница 11: ...side transfer of child into the seat Note Product codes 7556 7559 are provided with fixed armrests and therefore cannot be adjusted 3 1 4 Pelvic Strap When positioning concentrate on achieving a good...

Страница 12: ...Commode Chair Figure B Removal of the seat cushion reveals a commode seat Under the lid of the commode is a toileting bucket with lid see opposite The bucket has a handle to assist with transport for...

Страница 13: ...n Lieu Of Standard BK175 BK172 BK173 BK174 Adjustable Height Push Handle PH028 PH012 PH013 PH014 Fixed Height Push Handle PH029 PH006 PH007 PH008 Cushions 25mm Pelvic Cushions Pair CS005 CS006 CS007 C...

Страница 14: ...mits of the slots Ensure there is sufficient room for your client s feet at the front of the dolly 4 2 Buckle and Strap Operation 4 2 1 Buckle Strap Standard The buckle is secured by pushing K into L...

Страница 15: ...ootrest to desired setting Tighten M before use 4 4 Reversible Footboard The footboard and skis are attached to the chair by four screws provided The screws fit into the chair outriggers on each side...

Страница 16: ...re tight before use Always lock the castors when attaching items or making adjustments To lock apply foot to point A to unlock apply foot to point B 5 0 FINAL INSPECTION When the service is complete R...

Страница 17: ...t is not intended to be used in environments where there are rapid changes in the environmental temperature and humidity during intended use The Combi toileting chair suffers little from any effects o...

Страница 18: ...If cleaning is required after storage please see section 7 6 Cleaning of the user manual for instructions Shipping and storage conditions are also contained on Combi Box Label 6 3 Standards Applied Th...

Страница 19: ...hand wheels tighten correctly and are not cross threaded or damaged Check that the leg screws are present and are tightened correctly Check all straps for fraying and that buckles are not missing dama...

Страница 20: ...ill continue to provide full support beyond the nominal service life provided the following conditions are met 1 A full service schedule has been maintained 2 A full service and inspection is undertak...

Страница 21: ...cleaned by using a soft brush Do not soak or immerse the chair in water The pelvic strap can be cleaned with a damp cloth In cases of extreme soiling it can be machined washed at low temperature with...

Страница 22: ...mplete this record after each service repair inspection and or maintenance Photocopy additional pages as required 9 2 1 Product Information Model Size Date of Manufacture Serial Number 9 2 2 Service i...

Страница 23: ...will refund in full This warranty is in addition to your legal rights Goods will only be collected from the original delivery address A charge may be made where the goods cannot be repaired under the...

Страница 24: ...24 Smirthwaite Ltd 16 Wentworth Road Heathfield Newton Abbot Devon TQ12 6TL Part Number 999 1 07 Rev 2 Date Sep 21...

Отзывы: