Chapter 6
Gated Operation
71
Figure 42. Single Shot Result: Fluorescence Spot, 100 ns Width, 10 ns Delay, Binned Vertically
Procedure for Setting up and Performing a Swept Gate Experiment
(Variable Width, Variable Delay):
The sequence of steps for a Swept Gate experiment with variable width and delay is
similar to the Swept Gate experiment with fixed width and variable delay but with the
following exception:
In addition to entering differing start and end values for the Gate Delay, you would
also enter differing start and end values for the Gate Width.
Procedure for Setting up and Performing a Static Gate Experiment
(Fixed Width, Fixed Delay)
The sequence of steps for a Static Gate experiment is similar to the Swept Gate
experiment but with the following exceptions:
The Active Mode, selected in Step 100, would be
Repetitive
.
The Gate Width and Gate Delay values would be entered in the dialog shown below.
Note that the minimum gate delay from T0 to optical gate is ~10 ns:
Figure 43. Repetitive Gating Setup dialog box
Содержание PI-MAX 3 System
Страница 1: ...4411 0129 Version 1 A September 8 2010 4411 0069 ...
Страница 32: ...32 PI MAX 3 System Manual Version 1 A This page intentionally left blank ...
Страница 52: ...52 PI MAX 3 System Manual Version 1 A Figure 20 Safe Mode and Fast Mode Operation ...
Страница 60: ...60 PI MAX 3 System Manual Version 1 A Figure 24 Experiment is Master Clock Hardware Setup and Timing Diagram ...
Страница 72: ...72 PI MAX 3 System Manual Version 1 A This page intentionally left blank ...
Страница 90: ...90 PI MAX 3 System Manual Version 1 A This page intentionally left blank ...
Страница 100: ...100 PI MAX 3 System Manual Version 1 A This page intentionally left blank ...
Страница 110: ...110 PI MAX 3 System Manual Version 1 A Figure 68 Outline Drawing PI MAX3 with F mount Adapter ...
Страница 111: ...Appendix B Outline Drawings 111 Figure 69 Outline Drawing PI MAX3 with Spectroscopy mount Adapter ...
Страница 112: ...112 PI MAX 3 System Manual Version 1 A PI MAX3 Power Supply Figure 70 Outline Drawing PI MAX3 Power Supply ...
Страница 113: ...Appendix B Outline Drawings 113 CoolCUBEII Circulator Figure 71 Outline Drawing CoolCUBEII Circulator ...
Страница 114: ...114 PI MAX 3 System Manual Version 1 A This page intentionally left blank ...
Страница 120: ...120 PI MAX 3 System Manual Version 1 A This page intentionally left blank ...
Страница 140: ...140 PI MAX 3 System Manual Version 1 A This page intentionally left blank ...
Страница 142: ...142 PI MAX 3 System Manual Version 1 A ...