9.2.2
If the device does not appear in the connections window
The connection of the device to the
LaserDiagnosticsSoftware LDS may be
blocked by the firewall:
X
In
Windows > Control panel >
Firewall
, enable the UDP port 20034.
The UDP port should be enabled by a system administrator.
The network address of the PC is not
within the range of the device.
X
In Windows > Control panel > Net-
work and Sharing Center
, assign an
IP address to your PC that is in the
same address range as the PRIMES
device (e.g. 192.168.116.xyz).
The IP address of your PRIMES
device can be found on the identifica-
tion plate.
The IP address should be entered by a system administrator.
If several Network cards or a
USB3-to-Ethernet card are installed in
the PC, the connection between device
and LaserDiagnosticsSoftware LDS may
be blocked by the selection of the wrong
Network card.
1. Select the appropriate Network card
in the
Connections > All
window.
The device is displayed in the
Con-
nections
window
2. Click on the device.
3. Click on the
Connect to device
but-
ton.