XTR-PX2-C0006-IN004 Ceph option
2-5
2.7 Attaching EAR RODs
Insert each of the EAR RODs (PIN) into its EAR ROD HOLDER,
and secure it with the SET SCREW M4 4 FLAT (x3) (Screw bag A).
Match the height and position of both left and right EAR ROD.
EAR RODs
- During cephalometric exposure (LA), the patient should stand in front of the FPD facing the right.
- There are two EAR RODs. One is for the left ear and the other one is for the right ear.
For the right ear: Labeled with an X sticker. There is a ring inside the cone (See below)
For the left ear: Labeled with a Y sticker. There is a pin inside the cone (See below)
For the right ear
X
Ring
For the left ear
Y
Pin
SET SCREW M4 4 FLAT
Position the EAR RODs when attaching them.
SET SCREW
SET SCREW
Bracket
Ear Rod
Press the EAR
ROD into the
corner so that it is
secured
Direction of the patient in LA mode can be
changed on the Console.
See "Ear Rod direction" on System
Settings screen.
Initial calibration should be carried out by
the default conditon.