This manual is intended to assist the user in the safe and efficient operation of the Model DS-400 Digital Blood Pressure Monitor. The product must be used in accordance with the procedures contained 

in this manual and must not be used for purposes other than those described herein. It is important to read and understand the entire manual. In particular, please read carefully and become familiar with 

the section entitled  TIPS ON TAKING YOUR BLOOD PRESSURE.  

The user is cautioned that the system may fail to yield specified measurement accuracy if operated or stored in temperature or humidity conditions outside the limits stated in the specifications section of 

this manual. Also, an inaccurate reading may result if this instrument is used on a child's arm. Consult your physician if you wish to take a child's blood pressure.
Indications for Use
The DS-400 system is intended for the non-invasive measurement of systolic and diastolic blood pressure and determination of heart rate in adult patients, i.e., age 15 and above. The product is recommended 

for use by patients with labile blood pressure or known hypertension in a home care environment as an adjunct to medical management. The cuff will accommodate an upper arm circumference range of 

approximately 8.7 to 12.6 inches (22 to 32 centimeters). Pressure is measured over a range of 0 to 300 mmHg and heart rate over a range of 40 to 160 beats/minute.
Method of Measurement
This product employs the oscillometric method for measurement of blood pressure and heart rate. The cuff is connected to the main unit and wrapped around the arm. Circuits within the monitor sense the 

small oscillations in pressure against the cuff produced by the expansion and contraction of the arteries in the arm in response to each heart beat. The amplitude of each pressure wave is measured, 

converted to millimeters of mercury, and displayed on the LCD as a digital value. Memory circuit stores the 30 most recent measurement results for comparison.
Accuracy was evaluated during clinical studies in which results with the DS-400 system were compared with simultaneous auscultatory measurements obtained with a standard sphygmomanometer. In these 

studies, the onset of the fourth (or fifth) Korotkoff sound was taken as the diastolic pressure for the purpose of determining overall efficacy. A copy of this study is available from the manufacturer on request. 

To insure that accuracy is maintained, compare readings obtained with the DS-400 system with values measured by a trained observer using a manual sphygmomanometer at least every 6 months.

Blood pressure measurements determined with this device are equivalent to those obtained by a trained observer using the cuff stethoscope auscultation method, within the limits prescribed by the 

American National Standard, for Electronic or Automated Sphygmomanometers.

The DS-400 system has the ability to function according to its specifications in the presence of common arrhythmias such as atrial or ventricular premature beats or atrial fibrillation.

CAUTION! Do not use cuffs other than those manufactured for this product.

Do not use this instrument without consultation with your doctor if you are under dialysis therapy or on anticoagulants, antiplatelets or 

steroids. Use of this instrument under such conditions could cause internal bleeding.
To avoid any possibility of accidental strangulation, keep this unit away from children and do not drape AIR HOSE around your neck.
For specific information on your blood pressure, contact your physician. Never make any judgment on measurement results on your own.
Use of this device in areas near portable phones, microwave ovens or other devices with strong electromagnetic field may cause malfunctions.

Model

: DS-400

Operating Principle

: Oscillometric method

Indicator

: 11 digits LCD display

Pressure indicating Range : 0 to 300 mmHg (Cuff pressure)
Measuring Range

: 50 to 250 mmHg (systolic)

40 to 180 mmHg (diastolic)
40 to 160 bpm (pulse rate)

Accuracy

:

±

3 mmHg (Cuff pressure)

±

5% of reading (pulse rate) 

Inflation

: Squeeze bulb

Deflation

: Constant air release valve

Exhaust

: Manual rapid exhaust valve

Power Supply

: 1.5 volt "AA" (R6P/LR6) type battery

Power Consumption

: 0.06W (max.)

Memory

: 30 measurement results, computation of the average of 

stored Data

Operating environment : +50ûF to+104ûF

, 15 to 90% relative humidity

Storage environment 

: -4ûF to+122ûF

, 15 to 95% relative humidity

Cuff

: Coverage arm circumference ; 8.7 to 12.6 in. [22 to 32 cm]

Main Unit

: Weight ; Approx. XX oz. [XXX g] (without batteries)

Size ; XX x XX x XX in [XXX x XXX x XXX mm] (W x D x H)

Specifications are subject to change without notice due to improvements in performance.

SPECIFICATIONS

BASIC PRODUCT INFORMATION

PRECAUTIONS FOR USE

Digital Blood Pressure Monitor DS-400

CONTENTS

Page

BASIC PRODUCT INFORMATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
PRECAUTIONS FOR USE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
TIPS ON TAKING YOUR BLOOD PRESSURE . . . . . . . . . . . . . . . . . . . . . . 2
BATTERY INSTALLATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2

CORRECT MEASURING POSTURE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
WRAPPING THE CUFF . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2

Page

MEASUREMENT PROCEDURES . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
MEMORY FUNCTIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
CARE AND MAINTENANCE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
ERROR DISPLAYS AND TROUBLESHOOTING . . . . . . . . . . . . . . . . . . . . . 4

BLOOD PRESSURE RECORD . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4

INSTRUCTIONS

E

NGLISH

PARTS NAMES AND PRODUCT COMPONENTS

POWER BUTTON

DISPLAY

CUFF

AIR HOSE

MEMORY BUTTON

AIR CONNECTOR

AIR PLUG

INFLATION BULB

INSTRUCTION MANUAL

CONSTANT AIR RELEASE SIDE

CUFF RING

RAPID EXHAUST SIDE

AIR CONTROL VALVE

Prevention® is a registered trademark of

RODALE INC.
©2005 RODALE INC.
All Rights Reserved
Produced under license by Mark of Fitness, Inc.