Quick Hook Up Guide and
Patient Check List
Attach the electrodes
© 2017 Preventice Solutions or its affiliates. All rights reserved.
Preventice Services, LLC and Preventice Technologies, Inc. are wholly owned subsidiaries of Preventice Solutions, Inc.
Activating the monitor
2
Caution: U.S. Federal law restricts this device to sale by or on the order of a physician.
1.
Attach the
black
wire aligned with the left side of your neck, two finger
widths below the left collarbone.
2.
Attach the
red
wire below and to the left of the black wire, under the
breast and to the outer ribcage (bone part of the ribs).
3.
Attach the
white
wire aligned with the right side of your neck, two finger
widths below the right collarbone.
4.
Lastly, attach the
green
wire below and to the right of the white wire,
under the breast and to the outer ribcage (bone part of the ribs).
All leads should be put on
BEFORE
the plastic
PULL Tab
is removed from the
monitor. The order of lead placement is
important
and should be done in the
correct order every time the leads are attached.
1
2
3
4
1
Once all four leads are properly connected, sit down and remove the
PULL
Tab
from the monitor. Remain still for the
first 60 seconds
.
You will hear a ring tone to confirm the monitor is powered on correctly.
The monitor will automatically proceed to the “Monitoring” screen.
LEFT
RIGHT