2.2.3. Adjusting the Measurement Range of a HART
®
Transmitter
While the instrument is being connected, in the
DEVICE
INFO
tab are shown
information about instrument identification, such as TAG, manufacturer, description,
message, date, measuring range and damping filter. Some of these parameters can be
changed in STANDARD CONFIG.
In the
DEVICE
INFO
tab, the
MIN
and
MAX
fields indicate the measuring range of
the HART
®
transmitter. For PV (primary variable) equal to the MIN value, the transmitter
should generate 4 mA. For PV (primary variable) equal to the MAX value, the transmitter
should generate 20 mA. The maximum allowable range of the transmitter is shown just
above (
RANGE ...
). To edit the range of the transmitter, just change the MIN and MAX
values and press the
SAVE RANGE
button.
On this screen you can also edit the unit of the primary variable and the input filter
(damping).
Fig. 58
- Adjusting the measuring range of the HART
®
transmitter