background image

 

 

Page 10

www.prestigemedical.co.uk

 

[email protected] 

 

3: OPERATION 

SETUP 

To setup the pouch sealer plug the power cord into the power socket ensuring 
the ON/OFF switch is in the OFF position while you do this. Remove the roll 
holder from the unit by sliding it away from the user. Insert one or more rolls onto 
the roll holder and reinsert the holder back into its housing. 

 

Press the ON/OFF switch to the ON position, the LED will then turn green and 
begin to flash. The heating element will now begin to heat up. Once the LED 
stops flashing after approximately 3 – 4 minutes the pouch sealer will emit a peep 

and the heating element will be at its 
required temperature; the pouch sealer is 
now ready for use. 

 

Please Note - If the lever is activated before 
the pouch sealer has reached the required 
temperature, the LED will turn red and the 
buzzer will sound. To prevent this return 
the lever to the down position and wait 

until the pouch sealer has reached optimum temperature.  

SEALING 

Unroll and insert the polypropylene pouch material into the rear slot of the front 
tray, check that the clear plastic film is face up. Pull enough of the pouch material 
through to fully cover the instruments that are going to be sealed. Close the 
sealing bar by pulling the lever to the vertical position. Manually operate the 
cutting tool by moving the slider from one side to the other (Making sure the 
sheet has fully sealed before cutting). 

Содержание Pouch Sealer

Страница 1: ...Instructions for use Pouch Sealer...

Страница 2: ...cal co uk The Prestige Medical Customer Service Team is available to provide advice and assistance during normal office hours To avoid delays when making contact please have the unit s Model and Seria...

Страница 3: ...1 INTRODUCTION 4 FORWARD 5 PRESTIGE MEDICAL POUCH SEALER 5 SPECIFICATION 6 INSTRUMENTS 7 2 INSTALLATION 8 GENERAL CONDITIONS 8 WALL MOUNTING 9 3 OPERATION 10 SETUP 10 SEALING 10 4 USER MAINTENANCE 13...

Страница 4: ...oads are kept within a water tight and air tight seal for effective autoclaving and sterilization procedure They are suitable for mainly dentists beauticians podiatrists and in certain medical or labo...

Страница 5: ...with the operating conditions Such instructions are confidential and the customer is liable not to disclose any information to third parties Such operating instructions and its annexes shall not be c...

Страница 6: ...ions Width 470 mm Height 300 mm Depth 270 mm Packaging dimensions Width 480 mm Height 310 mm Depth 280 mm Reference standards The unit was designed in compliance with the following standards 73 23 CE...

Страница 7: ...otherwise switch the unit off Steady red light Sealing has finished where bar has been closed for 3 seconds Re open sealing bar by raising lever Red flashing light Sealing bar broken Replace the seali...

Страница 8: ...laboratory dentist or similar surgery where access is only allowed by authorised personnel It must not be installed Near a heat source In an area containing volatile or explosive substances Below or...

Страница 9: ...quired Repeat this process for the remaining hole to support the bracket Check the fixings are secure before mounting the unit It is advisable to use open hook expansion bolts to allow for easy instal...

Страница 10: ...and the heating element will be at its required temperature the pouch sealer is now ready for use Please Note If the lever is activated before the pouch sealer has reached the required temperature the...

Страница 11: ...off automatically to protect the sealing bag and the components of the pouch sealer After each sealing cycle the pouch sealer will check its temperature and if necessary it will activate the heater t...

Страница 12: ...uch sealer on and does not use if for at least 30 minutes it will go into ECO mode the sealing bar switches off to prevent damage and energy waste When it is in this mode the LED will flash green at a...

Страница 13: ...he maintenance must immediately report any anomalies scratches fraying etc Prestige Medical will accept no responsibility when procedures recommended by our service engineers are not heeded All mainte...

Страница 14: ...cted Replacing the blade Check and or replace the blade annually or if the blade has become blunt and cannot perform an easy cut DISPOSAL AND SCRAPPING Please dispose of its different parts in accorda...

Страница 15: ...to persons Do Not Disassemble the pouch sealer This should only be undertaken by an authorised service engineer Clean with abrasive or corrosive chemicals Immerse the pouch sealer in water Use the po...

Страница 16: ...s without notice Paper Please recycle Prestige Medical Limited East House Duttons Way Shadsworth Business Park Blackburn BB1 2QR England Registered in England Reg No 2826793 Vat No 633707835 Registere...

Отзывы: