The condenser assembly under the stage can be shifted up or down by means of 
the control knob (left hand side under stage) to effectively move the light beam 
closer to (up) or away from (down) the specimen under observation.

Finally,  the  dimmer  control  also  allows  the  user  to  adjust  the  intensity  of  the 
halogen light.    Vary these different methods of regulating the light beam on the 
specimen to obtain the most effective picture for your eyes.

As a general rule you will find that the higher the magnification you use, the more 
light is required and so all light controls will need to be increased.

After using your microscope, turn the dimmer control to the lowest setting and
turn off the power to prolong the life of the halogen bulb.   

                                                     

MAINTENANCE

Like other optical instruments, this microscope should be kept in a cool, shady and 
dry place, free from dust, fumes and moisture.    If not stored in its box, cover with 
a hood to protect from dust.

If there is any dirt settled on the lens, wipe it off gently with some lens paper.   If 
alcohol is used, be careful not to let it penetrate through the lenses and dissolve 
the gumming.    Any dust settled on the lenses can be blown away with an air 
blower or wiped off with a clean soft camel hair brush.  

Please  keep  the  original  packing  for  shipment  in  case  of  need  for  repair  or 
servicing.

IMPORTANT:    Use of Immersion Oil

The only objective requiring oil is the x100 objective.    However, once oil is used it 
must  be  wiped  off  the  objective  (and  possibly  the  slide)  with  a  lens  tissue 
otherwise it can ruin the lens.    The use of oil also can lead to contamination of the 
x40 objective if it is not cleaned off immediately.    Unfortunately, it is common for 
busy operators to forget to do this.

We  strongly  recommend  that  you  try  to  use  the  x40  or  x60  objectives  first, 
whenever possible, when attempting to examine stained smears (including those 
stained for viewing bacteria) before using the x100 objective.    You may also wish 
to try using the x16 eyepieces as well to increase the magnification without the use 
of oil.

Содержание Vetlab 125

Страница 1: ...plies co uk Vetlab Premiere 125 Biological Binocular Microscope USER MANUAL To obtain the best results from your microscope please read this manual before using WARNING To prevent fire or shock hazard...

Страница 2: ...Bacteriology Histology Pathology Medicament Chemistry research and clinical examination It is also ideal for use in veterinary laboratories schools and colleges CONFIGURATION Eyepiece Binocular head N...

Страница 3: ...10 40 S 100 S Oil Stage Double layer mechanical stage Stage size 140mm 140mm Moving range 75mm 50mm Condenser N A 1 25 Abbe condenser with iris diaphragm filter Focusing Coaxial coarse fine focusing a...

Страница 4: ...how it behaves and what result it produces 7 Attach power cord to the back of microscope and connect to power source Your microscope is now ready for use OPERATING INSTRUCTIONS Set light dimmer contro...

Страница 5: ...e will appear to be very close to the slide Because the 100x is an oil immersion objective a drop of immersion oil should be applied in the gap between the objective and the specimen The 100x oil imme...

Страница 6: ...d dry place free from dust fumes and moisture If not stored in its box cover with a hood to protect from dust If there is any dirt settled on the lens wipe it off gently with some lens paper If alcoho...

Страница 7: ...b Remove the old bulb Important Do not touch the halogen bulb with bare hands Wear gloves or hold the bulb using tissue or lens paper Body oils may damage the halogen bulb Hold the new bulb must be sa...

Страница 8: ...lting from abuse or misuse repairs or alterations performed by other than authorised repair technicians or damage occurring in transit If you have any problems with your Vetlab Premiere 125 microscope...

Отзывы: