OLT-301
S
ERIES
Optical Loss Tester
Operation Guide
V.1.23.17
Страница 1: ...OLT 301 SERIES Optical Loss Tester Operation Guide V 1 23 17...
Страница 2: ...2 3 Indicator 8 2 3 Use of Battery 8 2 4 Connector Cleaning 8 3 Operation 9 3 1 Power On 9 3 2 Stabilized Laser Source SLS 11 3 2 1 SLS Laser On Off 12 3 2 2 SLS Reference Setting 12 3 2 3 SLS Setup 1...
Страница 3: ...RL Test 22 3 5 2 ORL Save Record 23 3 5 3 ORL Recall 24 3 5 3 ORT Setup 24 4 Maintenance and Calibration 25 4 1 Cleaning of the Connectors 25 4 2 Calibration Requirement 25 5 Warranty Information 25 5...
Страница 4: ...voided laser beam may converge on the retina and cause permanent eye injury CAUTION Battery OLT 301 Series battery type is lithium battery Battery Power Supply Do not expose battery to fire or intense...
Страница 5: ...o the shipping agent or the representative of PRO immediately Retain the original packing materials in case reshipment is necessary 1 2 Unpacking and Inspection This instrument has been carefully pack...
Страница 6: ...o test records save in local unit remote unit both units USB power charging Over 35 hours continuous operation Multilanguage support No warm up quick start High resolution color LCD Pocketsize lightwe...
Страница 7: ...cal Input Output The OLT 301 Series is available with FC connector Interchangeable SC ST Note Output port is marked with LS Input port is marked with PM 2 USB Power Data Port The OLT 301 Series can be...
Страница 8: ...w the instructions below when cleaning Turn off the instrument before cleaning Non compliant operation may result in hazardous radiation exposure Turn off laser source before cleaning optical interfac...
Страница 9: ...Operation 3 1 Power On Press and loading screen appears Then it enters functional interface The last interface before last shutdown see Figure 3 2 Figure 3 2 The toggle between functional interfaces P...
Страница 10: ...ess to enter the corresponding function interface The toggle between different setting interfaces Under function interface press and to toggle the following interfaces Under function interface press a...
Страница 11: ...ess and to select Mass storage PC recognition it for a USB flash drive Communication the equipment directly transfer the data to the PC software mainly used in the optical power meter Draw Trance func...
Страница 12: ...etween dBm and mW 3 2 2 SLS Reference Setting Note This function is only for AutoID mode applicable Ref XXX XXdBm parameters will set up to the opposite side which has the corresponding models test in...
Страница 13: ...engths 3 2 3 SLS Setup Press and to Setup interface see Figure 3 7 Note This function is only for Auto ID mode applicable Figure 3 7 While the cursor on Auto ID press to select Manual or Automatic Und...
Страница 14: ...ce Precision Rated Optics laser source with AutoID function on the other side of the tested fiber enables AutoID the OPM will automatically set the reference value sent from the laser source as refere...
Страница 15: ...nterface see Figure 3 10 Figure 3 10 Press and to select item Del all Del current Go to Index Prev and Next 3 3 3 OPM Trace Press and to Trace sub interface see Figure 3 11 Figure 3 11 Press and to se...
Страница 16: ...nd 8hour Status press to running stop trace 3 3 4 OPM Setup Press and to Setup sub interface see Figure 3 13 Figure 3 13 Press and to select item Zero and Unit Zero Screw on the dust cap press the twi...
Страница 17: ...ue 5 Save 6 Toggle the wavelength and then repeat step 3 4 and 5 3 4 Optical Loss Test OLT Press and to Optical Loss Test interface see Figure 3 14 Figure 3 14 3 4 1 OLT Reference Setting Press and to...
Страница 18: ...easonable which is recommended After connecting the ports press and to select Start press to start the test Press and to select Save press to save the measured value as reference value 3 4 2 OLT Bi di...
Страница 19: ...gure 3 17 Figure 3 17 Press and to select item Cable and Fiber press to enable digit adjustment press to confirm Press and to select Save and press to save current record 3 4 4 OLT Recall Press and to...
Страница 20: ...lt PASS or FAIL according to the thresholds setting Loss dB km In bi directional test the unit will calculate the fiber s approximate length according to the attenuation dB km set here Please note tha...
Страница 21: ...ly calibrate zero 2 Reference Setting Ref Type set to twin 3 Threshold Setting set Thres and Loss the value of the Thres limit the maximum loss of the entire link Loss is generally provided by the fib...
Страница 22: ...Return Loss Test ORL Option Press to Optical Return Loss sub interface as shown in Figure 3 22 Figure 3 22 3 5 1 ORL ORL Test Press and to Test sub interface as shown in Figure 3 22 Press and to selec...
Страница 23: ...t return loss value as the ORL Zero ORL Ref 14 7dB 14 7dB means the current ORL value measured under default ORL zero value and default ORL reference Connect the reference patch cord to OLT port press...
Страница 24: ...sub interface as shown in Figure 3 24 Figure 3 24 Press and to select item Del all Del current Go to Index Prev and Next 3 5 3 ORT Setup Press and to Setup sub interface as shown in Figure 3 25 Figur...
Страница 25: ...Keep the cleanness of connectors and dust caps The detector needs to be cleaned timely 4 2 Calibration Requirement Calibration of OLT 301 Series is recommended every three years Please contact PRO Te...
Страница 26: ...n Materials Authorization number And describe briefly reasons for the return of the equipment to allow us offer you more efficient service NOTE To return the instrument in the case of repair calibrati...
Страница 27: ......
Страница 28: ...on Rated Optics Inc Corporate Office Billing Processing PO Box 877 Trexlertown PA 18087 Precision Rated Optics Inc Product Distribution Center Manufacturing Testing 9999 Hamilton Blvd Breinigsville PA...