Precision Medical EasyPulse 5 Скачать руководство пользователя страница 1

U S E R   M A N U A L

300 Held Drive 

Tel: (+001) 610-262-6090

Northampton, PA 18067 USA 

Fax: (+001) 610-262-6080

ISO 13485 Certified 

 www.precisionmedical.com

SAVE THESE INSTRUCTIONS

(198705 shown)

Federal (USA) law restricts this device to 

sale by or on the order of a physician.

1900 Series

(CGA 870 Connections)

Содержание EasyPulse 5

Страница 1: ...62 6090 Northampton PA 18067 USA Fax 001 610 262 6080 ISO 13485 Certified www precisionmedical com SAVE THESE INSTRUCTIONS 198705 shown Federal USA law restricts this device to sale by or on the order...

Страница 2: ...ARATION OF CONFORMITY 15 RECEIVING INSPECTION Remove the Precision Medical Inc EasyPulse5 Oxygen Conserving Regulator Conserver from the packaging and inspect for damage If there is any damage DO NOT...

Страница 3: ...minently hazardous situation which if not avoided will result in death or serious injury WARNING Indicates a potentially hazardous situation which if not avoided could result in death or serious injur...

Страница 4: ...Conserver Wash and dry hands properly prior to use DO NOT use a humidifier with the Conserver DO NOT allow cylinders to tip or fall Secure gas cylinders so they cannot fall For optimum safety keep cy...

Страница 5: ...nsure proper function and to avoid the risk of fire and burns Wind or strong draughts can adversely affect accurate delivery of oxygen therapy Using this equipment beside an open window or in front of...

Страница 6: ...ys The flow is determined by setting the flow control knob to the prescribed flow The oxygen is supplied to the patient through the cannula CAUTION Only personnel instructed and trained in its use sho...

Страница 7: ...uency Up to 35 b min Cannula Requirement Maximum 7 foot 2 1 m long standard adult single lumen nasal cannula Operating Conditions Temperature 35 F to 105 F 1 7 C to 40 6 C Altitude Sea level to 10 000...

Страница 8: ...OMPONENT DESCRIPTION CAUTION Missing or illegible labels must be replaced contact Precision Medical Inc Index Pins Yoke Seal Cannula Connection T Handle Dial Contents Gauge Indicator Pointer 198705 sh...

Страница 9: ...to tighten T handle this can lead to over tightening and will cause damage to the Conserver Be sure all connections are tight and leak free DO NOT use liquid leak detector to test for leaks Note For O...

Страница 10: ...quipment 2 Place the cannula over your ears and position the prongs in your nose as instructed by your Provider or cannula manufacturer 3 Slowly open cylinder valve counterclockwise until completely o...

Страница 11: ...kPa and below it is recommended to change to a full oxygen cylinder NO OXYGEN is delivered in between settings WARNING Use only Precision Medical Inc carry bag designed for the EasyPulse5 to prevent...

Страница 12: ...ackaged in sealed containers to prevent damage Precision Medical Inc will not be responsible for goods damaged in transit Refer to Precision Medical Inc Return Policy available on the Internet www pre...

Страница 13: ...in OFF position 3 Cylinder empty 4 Dial not set at proper position 5 Conserver not sensing breath 1 Open the valve on cylinder 2 Set to prescribed setting 3 Replace cylinder 4 Set dial so indicator po...

Страница 14: ...TS Description Part Cannula 504833 Yoke Seal 1221 Carry Bag with shoulder strap Conserver properly installed in Carry Bag ACCESSORIES Description Part Carry Bag M6 503920 Carry Bag M4 M6 504184 Carry...

Страница 15: ...make oral warranties about the merchandise described in this contract and any such statements shall not be relied upon and are not part of the contract for sale Thus this writing is a final complete a...

Страница 16: ...of Conformity see our website http www precisionmedical com documentation Precision Medical Inc 300 Held Drive Northampton PA 18067 USA Emergo Europe Prinsessegracht 20 2514 AP The Hague The Netherla...