Input Group (G.IN)
The Input Group includes parameters to select the
input type, define display units, and scale a process
input.
It also contains parameters related to the temperature
reference junction, input filtering, and making input
calibration adjustments.
It is recommended to
begin programming by
configuring the Input Group (G.IN)
.
Changing certain input parameters will reset
most other setup parameters.
Input Type (INP)
The Input Type (INP) parameter is for selecting the
input type. Refer
to Figure 23. Input Type and Range
Selection
to find the parameter setting display
selection for the desired input type.
Unit (UNIT)
UNIT can be set to °C and °F when a thermocouple or
RTD input is selected. The calculation to display each
unit type is done automatically.
For mA, VDC, or mVDC inputs, the units may be set
to °C and °F, %, or no unit (NONE). The selected unit
is displayed, but scaling must be done to properly
read the selected unit.
Decimal Point Position (DP-P)
This parameter sets the number of decimal places to
be used for mA, VDC, and mVDC inputs. It is used for
the display values as well as the scaling parameters
SL-L and SL-H.
Input Type Display
Range
K
K0
-328 to 2498°F
-200 to 1370°C
K1
-148 to 932°F
-100.0 to 500.0°C
J
J0
-328 to 2192°F
-200 to 1200°C
J1
-328 to 1652°F
-199.9 To 900.0°C
E
E1
-328 to 1652°F
-199.9 to 900.0°C
T
T1
-328 to 752°F
-199.9 to 400.0°C
R
R0
32 to 3092°F
0 to 1700°C
B
B0
212 to 3272°F
100 to 1800°C
S
S0
32 to 3092°F
0 to 1700°C
L
L1
-328 to 1652°F
-199.9 to 900.0°C
N
N0
-328 to 2372°F
-200 to 1300°C
U
U1
-328 to 752°F
-199.9 to 400.0°C
W
W0
32 to 4172°F
0 to 2300°C
PLII
PL0
32 to 2372°F
0 to 1300°C
JPt100
JPt0
-328 to 932°F
-200 to 500°C
JPt1
-328 to 932°F
-199.9 to 500.0°C
Pt100
Pt0
-328 to 1184°F
-200 to 640°C
Pt1
-328 to 1184°F
-199.9 to 640.0°C
4-20 mA
1-5V*
-1999 to 9999
1-5 VDC
1-5V
0-5 VDC
5V
0-10 VDC
10V
0-50 mVDC
0.05V
0-100 mVDC
0.1V
*Connect a 250 Ω resistor across the input terminals.