![PP Systems CIRAS-3 Скачать руководство пользователя страница 84](http://html1.mh-extra.com/html/pp-systems/ciras-3/ciras-3_operating-manual_1592668084.webp)
CIRAS-3 Operation Manual V. 1.09
84
Section 8. Operations
Operations (F1)
From the numeric or graph display, press
Operations (F1)
to access numerous system functions
including:
•
Recording Options (Rec Options – F2)
•
Set Clock (F3)
•
View saved data (View Saved – F4)
•
Calibration (F5)
•
Diagnostics (F6)
•
Firmware Upgrade (FW Upgrade-F7)
•
Help (F8)
Recording Options
Press
Operations (F1) > Rec Options (F2)
to set up a recording session. There are 3 options available
•
Manual recording – Normally used for individual leaf level gas exchange measurements.
•
Timed recording
•
Response curves – For automated, pre-programmed scripts for A/Ci curves, light
response, etc.
Manual Recording
Select the desired recording option at the top of the screen by using the Left or Right arrow keys – the
first option is
Manual recording
. For most leaf gas exchange measurements in the field, this method of
recording is most common.