CIRAS-3 Operation Manual V. 1.09
180
External Battery Pack (For extended operation in the field)
A convenient battery pack (STD551) can also be used for extended operation in the field.
Re-Order Information
Part Number
Description
STD551
Belt Pack Kit – Battery, Charger and Power Supply
Air Sampling Pumps (Reference and Analysis)
The CIRAS-3 has 2 rotary vane pumps used for sampling the reference and analysis air and should be
inspected periodically. Rotary vane pump components wear with prolonged use and material from the
vanes may build up over time within the pumps. The risk of this situation from occurring can be
minimized by regular pump cleaning with alcohol.
Typically, the following symptoms are attributed to a worn or faulty reference or analysis pump:
1.
Noise
– a worn pump usually sounds rough or vibrates.
2.
Temperature
– the outer casing (near the pump serial number) feels warm compared to other
pumps.
Please note.
We changed over to a new air sampling pump in July 2017. All instruments supplied prior
to July 2017 would have been supplied with Part Number 10127-1 (Manufacturer: Thomas) and all
systems supplied starting July 2017 would have been supplied with Part Number 10181-1 (Manufacturer:
Schwarzer).
Battery Pack (STD551)
Connection to CIRAS-3
via EXT PWR socket