Manual Kit Control Box with GUI
Page 9
Version: 2020-09-10
MANUAL
KIT CONTROL BOX WITH GUI
2.5 Calibration
The electronic calibration of the BiSS kit encoder is required to improve the measurement accuracy of the
kit encoder after installation. The calibration of the fully mounted sensor requires rotation of the motor shaft
at constant speed of 500 rpm, where deviation of angular velocity should be less than 2 rpm. For further
information, refer to Kit Encoder Manual.
Calibration Procedure
The calibration is performed by carrying out the following sequence:
1. Initiatethe procedure with the b
utton “Start Calibration” in the
main calibration window
Figure 6: Calibration main window
2. A pop-up window will give a signal to begin the rotation of the motor in CCW direction. The motor
should be run at constant rotational speed of 500 rpm < ±2 rpm. Once established, confirm with
“OK”
button.