NI TRONOX ® HD 

 

US ER’S  M AN UAL

 /  INS TRUCTIONS  

 

 

     

 
 
 

 

 

 
 
 

 

 
 

Содержание NITRONOX HD

Страница 1: ...www porterinstrument com nitronox Parker Hannifin Corporation Precision Fluidics Division Porter Instrument 245 Township Line Road Hatfield PA 19440 Office 215 723 4000 Fax 215 723 5106 NITRONOX HD USER S MANUAL INSTRUCTIONS ...

Страница 2: ... 44 0 1271 313131 Fax 44 0 1271 373636 The Quality System for Porter Instrument is certified to ISO 13485 The scope of our registration is The design manufacture distribution and servicing of fluidic components and integrated systems for use in the Medical Equipment Diagnostic Biopharmaceutical Agent Detection Analytical Instrumentation and Industrial markets Nitrous Oxide Oxygen Sedation Flowmete...

Страница 3: ...s Preparation of Mobile E Stand 8 Installations Preparation of Nitronox Demand Valve 9 Description 10 Functional Schematic 11 Specifications Functional Features Tests 12 Operation Maintenance 13 Functional Checks Tests 15 Cleaning 16 Troubleshooting 18 Warranty 19 Examination Examine shipping carton for signs of external damage Remove contents from carton and inspect for visible damage or missing ...

Страница 4: ...nd maintain the analgesia delivery system to prevent N2O leaks in all hoses connections and fittings Have all leaks repaired immediately WARNING Use Scavenging Medical workers are exposed to N2O during administration of N2O O2 conscious sedation analgesia Controls are effective in the patient treatment area to achieve low levels of ppm parts per million exposure Controls include System Maintenance...

Страница 5: ... prevent misconnection of Oxygen and Nitrous Oxide hoses DO NOT ATTEMPT TO CHANGE THE DIAMETERS OR CONNECTORS OF THE DEVICE Tampering with the cross protection system constitutes acceptance of liability by the installer Do not allow crossed lines to defeat the safety features of the Nitronox and or central gas supply manifold systems Crossed lines will create a dangerous and hazardous condition wh...

Страница 6: ...ination of medical professional The use with pediatric patients especially age 1 to 4 years requires caution and specific protocols developed by the medical professional upper age limit at determination of medical professional Weight limitations at the determination of medical professional Contraindications Inability to hold own face mask impaired oxygenation or hemodynamic instability Acute drug ...

Страница 7: ...istration mode over the course of several hours as secondary labor and end labor stages are experienced The upper limit of the number of hours of this described transient delivery is at the determination of the medical professional Patient Population Adult and Pediatric Used to deliver a gas mixture to a conscious spontaneously breathing patient who is awake alert and cooperative and requires reli...

Страница 8: ...ion of Nitronox nitrous oxide to pediatric patients typically ages 1 to 4 The concept that for certain circumstances the use of the Nitronox or the use of other demand valve delivery systems should have a protocol where the provider administers the nitrous oxide as opposed to complete self administration is in potential conflict with protocols for labor analgesia where there is an emphasis on self...

Страница 9: ...l start to flow Ceasing to inhale or if the patient stops breathing the valve automatically closes to stop the flow of the nitrous oxide oxygen mixture WARNING Do not use near open flame or in an unventilated area Can accelerate burning and be toxic Specifications Flow The demand valve through the Nitronox housing is to be connected to gas supply capable of delivering a minimum of 100 LPM 40 90 ps...

Страница 10: ... FM 916 for 4 Cylinder Mobile E Tank Stand Installation and Instructions FM 1330 for Medical Breathing Circuit Scavenger System User s Manual Instructions 4 Cylinder E Stand 2 Cylinder Mobile Cart With Nitronox With Nitronox 4 Cylinder E Stand 2 Cylinder Mobile Cart Wall Mount Mounting on Mobile Stand ...

Страница 11: ...o rings Fig 2 3 Tighten thumb screw of Retaining Bracket to the side of the Manifold to secure the Demand Valve Fig 3 4 Connect the Gas Delivery Hose to Demand Valve Fig 4 This is a quick connect attachment Push the two ends together until click sound is heard To remove push the metal clip in and pull apart 5 Attach Mask Hanging Bracket to the Left side of the Nitronox HD Fig 5 Fig 5 6 Assembled N...

Страница 12: ...ilable in various configurations offering three mounting styles 1 Nitronox with 4 Cylinder or 2 Cylinder Mobile Cart Fig 8 and cylinder mount 2 nitrous oxide and 2 oxygen E or D size Refer to FM 916 or FM 1346 for Installation and User Instructions 2 Nitronox with Mobile Stand only for use with pipeline gas supply only 3 Nitronox with wall mount for use with pipeline gas supply Use Nitronox in con...

Страница 13: ...11 Nitronox Inhalation Analgesia System 9 100cc min O2 bleed ...

Страница 14: ... 45 psi to 60 psi Fail Safe Check Out Test Initially and Every 6 Months NOTE To vent pressures Disconnect the Demand Valve from the Gas Delivery Hose by pushing the metal clip in and pull apart With Hose disconnected vent the system pressure by depressing the Hose check valve by pushing against it with a clean Phillips head screwdriver Verify both line pressure gauges are at 0 psi Note When testin...

Страница 15: ...nders Use Porter A 3399 000 replacement washers once yr or as needed Have spare washers 3 Warning Keep the cylinder top clean while performing attachment so as to minimize possibility of any foreign substance entering the regulator cavities minimize fire or explosion potential 4 Minimize leak risks E Stand With cylinder in position rotate swivel arm and move into secure locked position when Tee Ha...

Страница 16: ...served oxygen failsafe function may be quickly checked Remove the mask immediately and encourage mouth breathing With N2O supply pressure turned on turn off or disconnect O2 supply source pressure Disconnect the Demand Valve from the Gas Delivery Hose by pushing the metal clip in and pull apart Depress the Hose check valve which vents flow and O2 pressure will fall to 0 psi The flow through the Ho...

Страница 17: ...inder pressures Verify Nitronox line pressure gauges read within green band Note fail safe is open and pressure reaches Demand Valve Turn off N2O cylinder valve Verify that N2O E Stand gauge reads about 750 psi 50 bar Note exact needle position Verify that there is little or no movement up to increment of the exact needle position in 15 minutes Nitronox N2O line pressure gauge needle will stay in ...

Страница 18: ...mbly 6 Cleaning Methods We recommend the use of an approved disinfectant for the healthcare patient environment for cleaning the outside of the Nitronox device demand valve and accessories Do not spray disinfectant directly onto housing Spray disinfectant into disposable towel or use disinfectant surface wipe avoid excess disinfectant liquid and wipe unit thoroughly removing excess disinfectant to...

Страница 19: ... will not function properly Make sure that the flapper valve lies flat against its seat 5 If cleaning demand valve mounting bracket do not apply disinfectant directly on to O rings 6 Reassemble the parts Replacement Parts Outlet Adapter Housing Exhalation Valve Assembly Inspect the Demand Valve during the regular inspections of Nitronox Pages 15 16 Maintenance WARNING Do not disassemble or tamper ...

Страница 20: ...f 40 65 psi 2 8 4 5 bar Apparatus exposed to temperature below 32ºF 0ºC PRESSURE REGULATOR out of adjustment models with E cylinder yoke PRESSURE REGULATOR defective models with E cylinder yoke Allow apparatus to return to room temperature before making adjustment See Maintenance adjustment Step 1 Discontinue use and notify authorized distributor 5 Mixture pressures out of green band CONTROLLED PR...

Страница 21: ...y persons other than Parker Hannifin Corporation 3 not installed in strict compliance with applicable codes and ordinances or 4 not installed by an authorized Parker Hannifin Corporation dealer Parker Hannifin Corporation s obligation for breach of this warranty or for negligence or otherwise shall be strictly and exclusively limited to the repair or replacement of the product or part This warrant...