OPERATION

 

1. 

Set the mixture control to 100% Oxygen.

 

 

2. 

Set the Oxygen flow rate to equal total gas glow 
to be administered to the patient.

 

 

3. 

Apply nasal hood to patient.

 

 

4. 

Gradually decrease percentage of Oxygen until 
determined ratio is reached.

 

5. 

To remove patient from conscious sedation, set 
mixture valve to 100% Oxygen.

 

 

6. 

If Oxygen is required, press Oxygen flush 

button.

 

 

7. 

When procedure is completed, shut off 
machine.

 

 

8. 

If using a portable cylinder mount system, turn 
off the tanks.  If using a central gas system, 
disconnect the gas system then disconnect the 
gas supply hose from the outlet station.

 

 
 

 

 
MAINTENANCE AND SERVICE 

It  is  advisable, on  a  two (2) year cycle, to have the 
MXR Flowmeter factory checked and serviced.   
 
Perform “Functional Tests” Monthly as described on 
page 4. 
 

Inspect  and  maintain

  the  analgesia  delivery 

system  to  prevent  N

2

O  leaks  in  all  hoses, 

connections  and  fittings.    Repair  all  leaks 
immediately. 
 
All service and repair must be accomplished at 
Matrx by Parker.  Have your authorized distributor 
return the unit to our factory and arrange for a 
loaner unit for yourself. 
 

CLEANING METHODS 

We recommend the use of an approved disinfectant 
for  the  healthcare  environment  for  cleaning  the 
outside  of  the  flowmeter.  Do  not  spray  disinfectant 
directly  onto  meter.  Spray  disinfectant  into 
disposable towel and wipe unit thoroughly removing 
excess 

disinfectant 

to 

eliminate 

buildup.

 

Содержание Matrx MDM

Страница 1: ...ucts Division Mailing Address Riverside Road Pottington Business Park Barnstaple EX 31 1NP England Phone 44 0 1271 313131 Fax 44 0 1271 373636 The Quality System for Porter Instrument is certified to ISO 13485 The scope of our registration is The design manufacture distribution and servicing of Nitrous Oxide Oxygen Sedation Flowmeters Gas Scavenging Systems Gas Distribution Systems and Office Comm...

Страница 2: ...tection as well as that of the Doctor and the patients use 3 8 O D tubing for all Nitrous Oxide lines and O D tubing for all Oxygen lines To assure safe operation and conformation to local fire codes all Porter Instrument flowmeter systems are designed to be used with sedation delivery systems mounted inside walls and they meet or exceed the guidelines established by the National Fire Protection A...

Страница 3: ...Slide one end of breathing tube over outside diameter of breathing tube connector 3 Slide connector Y fitting of large nasal inhaler into free end of breathing tube Or N2O O2 Scavenger Large Scavenger and 3 Liter Breathing Bag Figure 2 Assembly Procedure 1 Position breathing bag as shown and slide opening in breathing bag over outside diameter of bag mount Figure 1 2 Slide the 22mm hose cuff of th...

Страница 4: ...ks at pressure connections on the unit Bubbles will appear at leaking locations when a soap water solution is used This procedure is recommended each time a cylinder is changed Refer to FM 916 Mobile E Tank Stand Installation and Instructions for connecting gas supply to the flowmeter and tank installation to mobile stand MDM SPECIFICATIONS Figure 3 Physical Size 5 wide x 9 high x 12 deep Weight 7...

Страница 5: ...MDM SPECIFICATION DIAGRAMS STANDARD MDM FLOWMETER HEADS Figure 3 4 ...

Страница 6: ...urn on manifold medical gas system as per manufacturer s instructions Adjust line pressures as instructed Pressure must be 50 55 PSIG 5 For portable systems utilizing E cylinder tanks refer to FM 916 Mobile E Tank Stand Installation and Instructions The Controls and Indicators Mixture Dial controls setting of oxygen concentration of gas mixture Adjustment of this control does not affect total flow...

Страница 7: ...ide flowmeter float falls at same rate as oxygen flowmeter float 5 Oxygen Flush Test Check Monthly Depress and release oxygen flush button to assure that breathing bag can be inflated 6 Non Rebreathing Valve Test Check Monthly Blow into the breathing tube connector Exhaled air must not pass through breathing connector 7 Air Inlet Valve Test Check Monthly Temporarily plug breathing bag mount and in...

Страница 8: ...tion MAINTENANCE AND SERVICE It is advisable on a two 2 year cycle to have the MXR Flowmeter factory checked and serviced Perform Functional Tests Monthly as described on page 4 Inspect and maintain the analgesia delivery system to prevent N2O leaks in all hoses connections and fittings Repair all leaks immediately All service and repair must be accomplished at Matrx by Parker Have your authorized...

Отзывы: