OPERATION
1.
Set the mixture control to 100% Oxygen.
2.
Set the Oxygen flow rate to equal total gas glow
to be administered to the patient.
3.
Apply nasal hood to patient.
4.
Gradually decrease percentage of Oxygen until
determined ratio is reached.
5.
To remove patient from conscious sedation, set
mixture valve to 100% Oxygen.
6.
If Oxygen is required, press Oxygen flush
button.
7.
When procedure is completed, shut off
machine.
8.
If using a portable cylinder mount system, turn
off the tanks. If using a central gas system,
disconnect the gas system then disconnect the
gas supply hose from the outlet station.
MAINTENANCE AND SERVICE
It is advisable, on a two (2) year cycle, to have the
MXR Flowmeter factory checked and serviced.
Perform “Functional Tests” Monthly as described on
page 4.
Inspect and maintain
the analgesia delivery
system to prevent N
2
O leaks in all hoses,
connections and fittings. Repair all leaks
immediately.
All service and repair must be accomplished at
Matrx by Parker. Have your authorized distributor
return the unit to our factory and arrange for a
loaner unit for yourself.
CLEANING METHODS
We recommend the use of an approved disinfectant
for the healthcare environment for cleaning the
outside of the flowmeter. Do not spray disinfectant
directly onto meter. Spray disinfectant into
disposable towel and wipe unit thoroughly removing
excess
disinfectant
to
eliminate
buildup.