SAFETY: 

 

ADA Guidelines 

In addition to the failsafe and other safety 
features found on most sedation machines, 
effective August 1976, the following 
specifications were added to those required for 
acceptance by the ADA Council on Dental 
Materials and Devices: 
 

1.  Emergency Air Valve 
2.  Rebreathing Check Valve 
3.  Resuscitator Quick Connect 

In addition, the ADA Council requires that: 

1.  The gas storage and delivery system 

meet the recommendations for the 
National Fire Protection Association.  
(See NFPA Code.) 

2.  The system be installed by a competent 

supplier of gases and equipment. 

 

NFPA Codes 

To assure safe operation and conformation to 
local fire codes, Porter Instrument Nitrous Oxide 
Sedation Systems meet or exceed the 
guidelines established by the National Fire 
Protection Association for Nonflammable 
Medical Gas Systems, NFPA 99.  Copies of 
NFPA 99 or portions thereof may be obtained by 
writing to 

National Fire Protection Association, 

Batterymarch Park, Quincy, MA  02269-9904  
USA

 or call:  1-800-344-3555 

 

 

INTENDED USE—Flowmeter for 
Analgesia Delivery System 

This device is intended for patient (adult and 
pediatric) use by an attending physician or 
dentist properly trained in its use.  Porter 
Instrument recommends the user be thoroughly 
familiar with the use of Nitrous Oxide - Oxygen 
Conscious Sedation for patient analgesia and be 
properly trained in its administration prior to 
using this product.  For training requirements on 
the administration of Nitrous Oxide - Oxygen 
Conscious Sedation, contact the appropriate 
regulatory authority in your country, state, or 
province.  Training is recommended to provide a 
practical, hands-on capability and an 

understanding of the behavioral aspects of 
Nitrous Oxide Sedation and will complement the 
safety features of this device. 

 

                                                  

 

 

EQUIPMENT PACKAGE 

FLEXIBLE INSTALLATION 

Item 

Part Number 

Description 

1 4065D 

Digital MXR-D Flowmeter 

2 PA-450-1-001 

O

2

 Hose, Green 20 Feet Long 

A-1679  or 

Remote Bag Tee Mount 

 A-2633-000 

Through Wall Adapter Assy for Bag Tee 

4 8005 

O

2

 DISS / DISS Hose, 5 Feet Long 

5 8505 

N

2

O DISS / DISS Hose, 5 Feet Long 

6 B-2460-001 

O

2

 DISS Shut-off Valve, with Check 

Valves 

7 B-2461-001 

N

2

O DISS Shut-off Valve, with Check 

Valves 

 

AVS-5000 

Automatic Vacuum Switch (Optional) 

 

A-2508-000 

Rubber Goods Hook (Optional) 

 

 
 

 

 

9

Содержание 4065D

Страница 1: ...DIGITAL MXR D Cabinet Mount Flowmeter Component Diagram and Installation Instructions MODEL 4065D FM 1062 Rev B 6 07...

Страница 2: ...CSA C22 2 No 601 1 M90 Symbol Description Protective Earth Ground Class I Device Alternating Current Earth Ground Attention consult the accompanying documents Type B Equipment Distribution of Porter M...

Страница 3: ...s 9 NFPA Codes 9 INTENDED USE Flowmeter for Analgesia Delivery System 9 INSTALL THE GAS LINES 10 SWITCHING THE UNIVERSAL CONNECTORS 10 CABINET WALL MOUNT INSTALLATION 11 FLOWMETER ACCESSORIES 13 CHECK...

Страница 4: ...is to be wired per NEC and local electrical codes The branch circuit breaker is the main disconnect Porter recommends the following 15A UL circuit breaker two wire with ground TANK REGULATOR SPECIFICA...

Страница 5: ...anual the safety protection provided by the equipment might be impaired The Porter Digital MXR D is designed to be used in normal environmental conditions 1 Indoor use 2 Altitude up to 8000 feet 2 438...

Страница 6: ...904 or call 1 800 344 3555 WARNING New or modified installations properly connected gas pipelines are absolutely essential to patient safety The dealer or contractor should provide written documentati...

Страница 7: ...nd alarms needed for the Vanguard test Then turn off both Oxygen tanks and leave both Nitrous Oxide tanks on For the Sentinel only one tank of each gas is on at a time Turn off the one open Oxygen tan...

Страница 8: ...outlet station the Oxygen 1 2 inch pipe is incorrectly decreased down to the DISS 3 8 inch size reserved for Nitrous Oxide lines Oxygen is fed into the Nitrous Oxide lines At gas outlet station the c...

Страница 9: ...Analgesia Delivery System This device is intended for patient adult and pediatric use by an attending physician or dentist properly trained in its use Porter Instrument recommends the user be thorough...

Страница 10: ...ter The dental dealer is responsible for the final connections to the flowmeter the details of flowmeter installation checking for crossed lines leak testing of connections to the flowmeter and basic...

Страница 11: ...ce a piece of tape over the rear power supply connector temporarily so dry wall dust does not enter the connector Wrenches required 7 8 for N2O connector 5 8 for O2 connector and 7 16 for mixed gas co...

Страница 12: ...e O2 shut off valve and the N2O DISS hose blue to the N2O shut off valve 9 Run 20 feet mixed gas hose from flowmeter mixed gas outlet hose barb to hose barb of Remote Surface Mount or Through Wall Ada...

Страница 13: ...ff position Confirm that the DISS Shut Off Valves are in the open position Pressurize the sedation gas supply lines with 50 PSI Observe any pressure decay after an overnight time period 5 PSI drop all...

Страница 14: ...POWER SUPPLY SYSTEM ILLUSTRATION 14...

Страница 15: ...GAS DISTRIBUTION SYSTEM ILLUSTRATION 15...

Страница 16: ...w Knobs to full mechanical stops Installation Check for proper display STEP 1 N2O L M There is no Oxygen flowing There is no Nitrous flowing N2O and Total Flow valves are shut tightly Normal display c...

Отзывы: