A-5
A TMS-A-305 Calibration Set
A.4 Evaluating the Measurement Result
The measurement is analyzed using the
CalibrationDataEvaluation
add-in.
The Add-in queries successively the used standard, and the ambient
temperature as well as the device to be calibrated. You can avoid these
queries by specifying the information in the file name of the measurement file
that is to be analyzed. For example the file
TMS-350_EN18_3_22°C.tms
contains all information required.
The order of the information in the file name is arbitrary. The names of the
device and the standard have to be given in the same way they have been
entered in the dialog. When entering the temperature you have to pay
attention that the characters
°C
are typed directly and without a blank after
the temperature value.
Execute Add-in
To analyze the measurement result, proceed as follows:
1. Select
Tool > Evaluate calibration data
.
The
Select Standard
dialog appears.
Figure A.5:
Select Standard
dialog
2. If you have defined several standards (refer to S
A.2), then select
the standard used and click
OK
. You will find the identification number of
your depth setting standard in the calibration certificate.
The
Temperature
dialog appears.