6-1
6 Technical Specifications
6 Technical Specifications
6.1
Harmonized Standards Applied
INFORMATION
The light source complies with risk category 1. Pay attention to the
information about lamp safety in S
INFORMATION
This device is subjected to Limit Class A and can cause radio interferences
with electrical devices in residential areas.
Take appropriate protective measures to exclude any such impairments.
6.2
TMS-E-350 Controller
Mains Connection
Ambient Conditions
Housing
Photobiological safety:
IEC / EN 62471:2009-03
(Photobiological safety of lamps and lamp systems)
Risk category:
1
Electrical safety:
IEC / EN 61010-1:2011-07
(Safety requirements for electrical equipment for
measurement, control, and laboratory use)
EMC:
IEC / EN 61326-1:2006-10
(EMC requirements on Emission and Immunity -
Electrical equipment for measurement, control, and
laboratory use)
Emission:
Limit Class A
IEC / EN 61000-3-2 and 61000-3-3
Immunity:
IEC / EN 61000-4-2 to 61000-4-6
and IEC / EN 61000-4-11
Mains voltage:
100 … 240 VAC ± 10%, 50 / 60 Hz
Power consumption:
max. 40 VA
Fuses:
2.0 A / slow-blow
Protection class:
1 (protective grounding)
Operating temperature:
+5 °C…+40 °C (41 °F…104 °F)
Storage temperature:
−
10 °C…+65 °C (14 °F…149 °F)
Relative humidity:
max. 80%, non-condensing
Dimensions:
240 mm x 140 mm x 420 mm
Weight:
5.5 kg