![Polytec PDV-100 Скачать руководство пользователя страница 37](http://html1.mh-extra.com/html/polytec/pdv-100/pdv-100_user-manual_1578100037.webp)
7-3
7 Technical Specifications
Measurement Range
Signal Output DIGITAL OUT
Digital output for the velocity signal
Low Pass Filter
Measurement
range
1
full scale
(peak)
1
adjustable via the display
Scaling factor
(analog output)
Resolution
2
2
The resolution is defined as the signal amplitude (rms) at which the signal-to-noise ratio is 0 dB in a
1 Hz spectral bandwidth (RBW), measured on 3M Scotchlite Tape® (reflective film).
Maximum
acceleration
mm / s
m / s
2
20
5
< 0.02
2 760
100
25
< 0.02
13 800
500
125
< 0.1
69 000
Data format:
S / P-DIF, 24 bit, 48 kSa / s
Frequency range:
0 Hz ... 22 kHz
Propagation delay:
approx. 1 ms
Calibration accuracy:
± 0.2% (0.05 Hz ... 22 kHz)
Filter type:
digital, FIR type
Cutoff frequencies ( 0.1 dB):
1 kHz, 5 kHz, 22 kHz (adjustable via the display)
Frequency roll-off:
> 120 dB / dec
Stop band attenuation:
> 100 dB
Frequency response:
see the following table and
Filter setting
Pass band ± 0.1 dB
Stop band
(attenuation > 40 dB)
1 kHz
0.5 Hz ... 1 kHz
> 4.3 kHz
5 kHz
0.5 Hz ... 5 kHz
> 8.4 kHz
22 kHz
0.5 Hz ... 22 kHz
> 25 kHz