E-1
E Interface Operation RS-232 and IEEE-488 / GPIB
Appendix E: Interface Operation RS-232 and IEEE-488 / GPIB
E.1 Configuration of the Interfaces
Using the interfaces RS-232 and IEEE-488 / GPIB the following commands
are supported:
•
query and change the settings of the controller
•
query the overrange
•
query the optical signal level
•
focus the laser beam only for vibrometers with the sensor head OFV-303
The interfaces can be configured via the display of the controller in the menu
CONFIG (refer to
).
RS-232
The two following configurations are possible:
•
9,600 Baud, 8 data bits, 1 stop bit, no parity bit (factory setting)
•
4,800 Baud, 8 data bits, 1 stop bit, no parity bit
IEEE-488 /
GPIB
For the interface IEEE-488 / GPIB, every address in the range from 0 up to
and including 30 can be set.
The factory setting is: Address = 5
The interface has the following functions which comply with the standard
IEC-488.1:
SH1, AH1, T4, L2, TE0, LE0, SR0, RL1, PP0, DC1, DT0, C0
E.2 Interface Commands
E.2.1 Control Commands RS-232
REM
Queries the operation mode:
GTL
Sets the status LOCAL (REM = 0).
REN
Sets the status REMOTE (REM = 1).
LLO
Sets the status LLO (LOCAL LOCK OUT) (REM = 2).
Answer
Answer with ECHOON
0 = LOCAL
REM 0
1 = REMOTE
REM 1
2 = LLO
REM 2