110-511 PSC/EN
11
Set-Up Sub-Menus
Press the
key to access the Temperature Controller’s set-up sub-menus. The
or
keys are used
to change the current setting / value in the sub-menu.
Sub-Menu
Selection / Range
To Change
Current Value
Temperature unit
°C or °F
Press
for °C
Press
for °F
Offset Calibration
-3.0°C to +3.0°C
Press
or
Low Limit
-35° to +10°C / -30° to +50°F
Press
or
High Limit
+40° to +140°C / +100 to +285°F
Press
or
To accept a value in a sub-menu, press
or allow the display to ‘time out’ (approximately 10 seconds).
Adjusting the Temperature Set Point
This is the temperature at which the fluid in your circulating bath will be maintained. It may be set to one-
tenth of a degree over a range of -50° to +135°C / -60° to +275°F. The factory default is 20°C / 68°F.
To Change:
Press the
or
key. The word “SET” will begin flashing. Continue holding
or repeatedly press
or
until the desired set point temperature is displayed.
To Accept:
Press
or allow the LCD to return to the main operational display
(approximately 10 seconds).
Flashes when set point
is being changed