110-874

Safety Recommendations

To prevent injury to personnel and/or damage to property, 

follow your workplace’s safety procedures when operating 

this equipment. You should comply with the following safety 

recommendations:

WARNING

:

• 

This Immersion Circulator is suitable only for use 

with Class I non-flammable fluids 

(per DIN 12876-1).

• 

Be aware of the chemical hazards that may be 

associated with the bath fluid used. Observe all 

safety warnings for the fluids used as well as those 

contained in the material safety data sheet.

• 

Use only recommended bath fluids.

• 

Use only non-acid bath fluids.

• 

Use fluids that satisfy safety, health and equipment 

compatibility requirements.

!

WARNING

:

• 

Connect the power cord on this unit to a properly 

grounded power outlet. Make certain that the outlet 

is the same voltage and frequency as your unit.

• 

Never operate the unit with a damaged power cord.

• 

The use of an extension cord is not recommended.

• 

Turn the unit OFF and disconnect mains power 

before performing any maintenance or service.

WARNING

:

• 

Never operate the Circulator without bath fluid in 

the reservoir. Periodically check the reservoir to 

ensure that the liquid depth is within acceptable 

levels. Refill the reservoir using the same bath 

fluid that is already in the reservoir. Bath oil must 

not contain any water contaminants and should be 

preheated to the actual bath temperature before 

adding as there is an explosion hazard at high 

temperatures.

• 

Drain all fluid from the reservoir before moving 

or lifting. Be sure to follow your organization’s 

procedures and practices regarding the safe lifting 

and relocation of heavy objects.

• 

Turn electrical power to the unit OFF and remove 

the Immersion Circulator from reservoir before 

draining.

!

Содержание LX

Страница 1: ...110 874 PSC EN 25 February 2014 LX Immersion Circulator Operator s Manual ...

Страница 2: ......

Страница 3: ...N 10 Main Operational Displays 11 Set Up Sub Menus 12 Adjusting the Temperature Set Point 13 Selecting the Temperature Unit 13 Setting and Using the Timer 14 Loss of Power Restart 14 Alarm Messages 15 Routine Maintenance and Troubleshooting 15 Maintaining Clear Bath Water 15 Cleaning Your Immersion Circulator 16 Resetting the Over Temperature Safety 17 Troubleshooting Chart 17 Technical Informatio...

Страница 4: ...or the precise temperature control of suitable liquids in a reservoir Here are some of the features of the LX Immersion Circulator Simple intuitive operation Displays actual and set point temperature simultaneously Powerful pump Attaches securely to both flat and rounded tank walls Converts any tank or vessel up to 20 liters into a reliable circulating bath Fully enclosed housing prevents direct c...

Страница 5: ...t read this manual thoroughly prior to working with the unit General Safety Information Read all instructions pertaining to safety set up and operation Proper operation and maintenance is the user s responsibility This symbol advises of danger from electricity or electric shock This symbol indicates a protective conductor terminal This symbol marks information that is particularly important This s...

Страница 6: ...n that the outlet is the same voltage and frequency as your unit Never operate the unit with a damaged power cord The use of an extension cord is not recommended Turn the unit OFF and disconnect mains power before performing any maintenance or service WARNING Never operate the Circulator without bath fluid in the reservoir Periodically check the reservoir to ensure that the liquid depth is within ...

Страница 7: ...te bath Regulatory Compliance and Testing The following are the compliance certifications for the equipment 120V Units ETL Certification UL 61010 1 CSA C22 2 No 61010 1 2nd Edition 07 12 04 R 10 28 2008 IEC 61010 2 010 2003 CSA C22 2 No 61010 2 010 2004 07 01 Ed 2 R2009 IEC 61010 2 051 2003 CSA C22 2 No 61010 2 051 04 Ed 2 R2009 240V Units CE Compliance Low Voltage Directive 2006 95 EC EN 61010 1 ...

Страница 8: ...firm proper operation In the unlikely event that the unit was damaged or does not operate properly immediately contact the company where your Immersion Circulator was purchased LX Immersion Circulator Quick Start Guide Operator s Manual Selection Guide Package Contents This product has been tested to the requirements of CAN CSA C22 2 No 61010 1 second edition including Amendment 1 or a later versi...

Страница 9: ...ponents Backlit Display Run Button Stop Button Increase Value Button Decrease Value Button Menu Button Cover Plate for Over Temperature Protection Reset Button Minimum Liquid Level Line Front Flow Port Maximum Liquid Level Line ...

Страница 10: ...8 110 874 Mounting Clamp Rear Flow Port Power Switch Electrical Power Cord Removable Heater Pump Access Cover Circulating Heating Actual Fluid Temperature Set Point Temperature Countdown Timer ...

Страница 11: ...NG If the proper fluid level is not maintained the heater coil may become exposed and possibly damaged fluid level too low or the bath may overflow fluid level too high 3 Plug the power cord into a properly grounded electrical outlet The LCD will light and the actual and set point temperatures will appear on the display WARNING Make certain that the electrical outlet is the same voltage and freque...

Страница 12: ...menus The items in these sub menus are used to select the temperature unit adjust the set point and set the timer Increase Value Increase the temperature set point or timer setting Decrease Value Decrease temperature set point or timer setting Turning Your Circulator ON Press the button on the rear of the unit All characters symbols on the LCD will momentarily light after which the actual and set ...

Страница 13: ... Circulating Heating Actual Bath Temperature Set Point Temperature Set Point Temperature Countdown Timer alternating display Standby Mode Run Mode Actual Bath Temperature May be shown if set temperature is greater than actual temperature ...

Страница 14: ... Menu button to access the Temperature Controller s set up sub menus To accept a value in a sub menu press the button Sub Menu Selection Range To Change Current Value Temperature Set Point 5 to 98 C 41 to 208 F Press or until the desired temperature value is displayed Temperature Unit C F Press once for C Press three times for F Timer 00 05 to 99 00 Press twice and then enter the desired time peri...

Страница 15: ...tory default is 20 C 68 F 1 Press once for C or three times for F 2 Press and or until the desired set point temperature is displayed 3 Press The temperature units sub menu allows you to select the temperature unit in which the actual bath temperature and set point temperature are displayed The factory default is C 2 Press 1 Press once for C or three times for F Current Temperature Unit Selecting ...

Страница 16: ...has elapsed the Circulator will beep three times go silent for 15 20 seconds and then repeat The Circulator will continue circulating the bath fluid and maintaining it at the set point temperature until the Stop button is pressed this also resets the timer 3 Press 1 Press twice Loss of Power Restart If electrical power lost while your Circulator is in use it will begin operating automatically once...

Страница 17: ...ation s procedures and practices regarding the safe lifting and relocation of heavy objects WARNING To avoid the potential for burns allow the unit to return to ambient temperature before cleaning or performing any maintenance WARNING Always turn your Circulator OFF and disconnect it from the electrical power outlet before performing any maintenance or service CAUTION Always clean and dry your LX ...

Страница 18: ...perature Controller or allow them to enter the Controller s vents Do not use abrasives as these could scratch the housing or the digital display CAUTION Never immerse the Circulator s controls or display in water or other liquids or place under running water Do not clean the unit in a glassware washer Bath Reservoir and Wetted Components A vinegar solution or calcium lime rust remover can be used ...

Страница 19: ...n Standby mode Insufficient fluid in reservoir Pump impeller jammed Press Run button Add fluid to reservoir Inspect pump and remove debris as required Over Temperature Safety Reset Button The Circulator s automatic over temperature safety removes power from the unit whenever the fluid temperature exceeds a factory set temperature This safety is reset as follows 1 Place the power switch on the rear...

Страница 20: ...Run button Add fluid to reservoir Increase temperature set point Insufficient Heating Insufficient circulation Low line voltage Ambient temperature too cool Excessive heat loss See Insufficient circulation above Check and correct as required Increase ambient temperature or relocate unit Check for vapor heat loss from internal reservoir Temperature unstable Insufficient circulation Debris or minera...

Страница 21: ...F 1200 watts 120V 60Hz 10 amps or 240V 50Hz 6 amps Indoor use only Maximum Altitude Operating Ambient Relative Humidity Installation Category Pollution Degree Ingress Protection Output Waveform 2000 meters 5 to40 C 41 to104 F 85 non condensing II 2 IP X0 Sinusoidal Reservoir Fluids A variety of fluids can be used with your Immersion Circulator The bath fluid must be chemically compatible with the ...

Страница 22: ...quipment Disposal WEEE Directive This equipment is marked with the crossed out wheeled bin symbol to indicate it is covered by the Waste Electrical and Electronic Equipment WEEE Directive and is not to be disposed of as unsorted municipal waste Products marked with this symbol must be collected separately according to the regulatory guidelines in your area It is your responsibility to correctly di...

Страница 23: ...mber is assigned and the product should be shipped transportation charges pre paid in either its original packaging or packaging affording an equal degree of protection to the indicated service center To insure prompt handling the return authorization number must be placed on the outside of the package A detailed explanation of the defect should be enclosed with the item The warranty shall not app...

Страница 24: ...22 110 874 Manufactured by PolyScience 6600 W Touhy Avenue Niles IL 60714 USA 1 800 229 7569 1 847 647 0611 www polyscience com ...

Отзывы: