37
Instructions for Use 2019-03-20
Troubleshooting
If the device does not seem to be working properly, the user should first try to inspect and adjust it himself.
* If the user is unable to inspect the device himself or if the device fails to work properly after being adjusted or after parts are replaced, contact
Poldent Co. Ltd.
Problem
Check Points
Response
Does not turn on.
• Check battery power.
• Charge battery
• Check battery installation.
• Install battery properly.
No beeping sound.
• Check if the Beeper Volume is set to “OFF”.
• Set it to “Low” or “High”.
Beep sounds even when
device is not being used.
• Device may be set for reverse rotation.
• A beep sounds periodically whenever the device is set
for reverse rotation. Turn the beeper off if it is annoy-
ing. (This will stop all beeping except when the device is
turned on.)
Backlight color does not
change.
• See if this function has been turned off.
• Turn this function on, if necessary.
Motor does not start
when file is inside canal.
• Is Endostar Navigator properly connected and
turned on?
• Check transmission cable connections. Turn on the
Endostar Navigator.
• Is the contrary electrode for the Endostar Naviga-
tor hooked in the patient’s mouth?
• Hook the contrary electrode in the corner of the pa-
tient’s mouth.
• Is the Apical Action Function set to “OFF”?
• Set it to “ON”.
• Is the Auto Start & Stop set to “OFF”?
• Set it to “ON”.
• Has the meter gone past the Flash Bar?
• Set the Apical Reverse or Stop to “Rev” (reverse).
Motor starts but then
stops right away.
• Did you hold down the Main Switch for more
than 1 second?
• If you hold the Main Switch down for more than 1 sec-
ond, the motor runs only while the switch is held down
and stops when it is released. The motor will run without
stopping if you release the switch in less than 1 second.
• Does “Abn.Stop LowBat” appear in the display?
• Very low battery power. Charge battery.
Motor reverses rotation
on its own.
• Check the Reverse Torque setting.
• The Torque Reverse can be turned off with “TRL” set-
ting.
• Check if the Apical Reverse or Stop is set to “Rev
(reverse).
• You can change it to “Stp” (stop).
Motor reverses rotation
too quickly.
• Check the Reverse Torque setting.
• Increase its setting value.
• Is the Apical Torque Reduction set to “ON”?
• The torque reverse value goes down as the file ap-
proaches the apex if the Apical Torque Reduction is set
to “ON”. Set it to “OFF” to keep the torque reverse value
constant.
Motor runs back and
forth continuously.
• Is rotation mode set to “OTR”?
• Torque load is greater than the setting for OTR mode.
• Does it do this even after calibration?
• Increase the setting for the Reverse Torque by 1 level.