Page 37/68
4.1.6 Export data (only plusoptiX A12C)
All patient data and measurement results are stored in the database. This database can be
used to create backup copies and reports.
Note
:
To save storage space, video files are not stored in the database. The last image
for each measurement is stored together with the measurement report (see
chapter 4.1.9 "Save and print a measurement report") on the SD card. In this
manner, the database can contain up to 100,000 entries (patients or
measurements).
Touch Export button (1) in Settings to move to this screen:
To prevent data loss, you should save a copy of your database on a regular basis. To do so,
connect an USB storage device and ensure that no filters are activated (i.e. none of the
buttons on screen are highlighted in orange). Now touch Checkmark button (2). The entire
patient database is now saved on the USB storage device as a DB file.
Note
:
Checkmark remains inactive (3) until an USB storage device is connected.
This backup copy can be used to restore a database. To do so, connect an USB storage
device containing the DB file in its root directory to the device. A prompt appears as to whether
the saved database should be restored. Where multiple databases exist in the root directory of
the USB storage device, the most recent (by save date) will be restored after confirmation.
1)
2)
3)
Screenshot 21: Saving a backup copy