6.3 Rinse Time
The RINSE function is used to keep pumps and tubes wet and to prevent salt crystal formation
which could lead to clogging of the fluid path and/or damage the pump.
By default, the RINSE function will execute after the IntelliPlex 1000 πCode Processor was idle for
2 hours
,
and as long as the instrument remains idle, the system will perform RINSE function
repeatedly. The RINSE function will rinse the tubes with ddH
2
O. Once the RINSE function is
completed, a pop-up window will appear on the screen instructing to start the PRIME function,
which will prime the tubing with Wash Buffer. The PRIME function can be performed immediately
after the RINSE function or later but must be completed before performing the next run. Please
make sure the wash buffer and waste bottle fill levels are adequate before running PRIME
.
To adjust the idle time triggering the RINSE function follow the steps below:
Step 1.
Open the System Menu
Step 2.
Select Setting on the menu page.
Step 3.
Select Rinse Time in the setting menu.
Step 4.
Set up the Rinse Time. The automated rinse can be set between 10 minutes or 24 hours.
Note: We recommend keeping the default settings. Please keep in mind that the instrument will
perform another RINSE cycle if the instrument remains idle for 2 hours after the PRIME function
was executed.