40
Section 5.2: Command Details
Serial command:
A
Description: Attention and put IBEAM in REMOTE mode.
Usage:
A<cr>
Example:
Send “A<cr>”. The IBEAM will switch to REMOTE mode (if not already in
REMOTE mode).
Response:
<lf><cr>OK<lf><cr><eot>
Serial command:
AB
Description: Set the accelerator to beam (A/B) ratio.
Usage:
AB(0-99)<cr>
Example:
Send “AB10<cr>”. The A/B ratio will be set to 10%. If the accelerator current
exceeds 10% of the beam current, an error will be displayed. Acceptable range is
from 0 to 99%. Default is 10.
Response:
<lf><cr>OK<lf><cr><eot>
Serial command:
AC
Description: Enable Auto Cathode Mode (in Remote mode). When enabled, the cathode filament
current will increase automatically until one of two conditions are met; either the
cathode current limit, or the discharge current threshold setting.
Usage:
AC(0 or 1)<cr>
Example:
Send “AC1<cr>”. Auto Cathode mode will be enabled. Sending “AC0<cr>” will
disable Auto Cathode mode.
Response:
<lf><cr>OK<lf><cr><eot>
Serial command:
AV
Description: Set the accelerator voltage.
Usage:
AV(0-1000)<cr>
Example:
Send “AV250<cr>”. The accelerator voltage will be set to 250 V. Acceptable range
is from 0 to 1000 V. Below 100 V is not recommended.
Response:
<lf><cr>OK<lf><cr><eot>