8.4 Exposure switch
The exposure switch can be mounted on the wall, or it can be hung from the
hook provided on the column top if a protected area is within reach.
A green light flashes on the exposure button when the X-ray system is
getting ready for an exposure. The green light stops flashing and stays on
continuously when the X-ray system is ready for an exposure.
During exposure a yellow radiation warning light illuminates on the exposure
switch. It indicates that the X-ray unit is generating radiation.
1 Exposure switch
2 Exposure button
8.5 Emergency stop button
Press the emergency stop button to stop the X-ray unit operating in an
emergency. All movements of the X-ray unit will be blocked and no radiation
will be generated.
A help message will appear on the control panel. Guide the patient away
from the X-ray unit. Then release the emergency stop button. The X-ray unit
will automatically restart.