17 Service
The X-ray unit must be checked by a qualified Planmeca service technician
once a year or after every 10 000 exposures (if this is sooner). This will
guarantee patient and user safety and ensure consistent image quality.
The annual maintenance service includes inspection of the following:
•
X-ray unit adjustments and quality control checks
•
Data security updates
•
Exposure switch
•
Exposure indicator lights and warning signals
•
Emergency stop button
•
Column motor nut
•
Labels
17 Service
User's manual
Planmeca Viso 101
Содержание Viso
Страница 1: ...Planmeca Viso user s manual EN 30005931 30005931...
Страница 10: ...Warning hot surface Standard ISO 7010 General warning Standard ISO 7010 4 Symbols 6 Planmeca Viso User s manual...
Страница 45: ...Use the segmentation buttons to select the active exposure areas 10 2D exposure User s manual Planmeca Viso 41...
Страница 79: ...Results The image is shown on the computer screen 11 3D exposure User s manual Planmeca Viso 75...
Страница 102: ...Patient supports 16 Cleaning and disinfection 98 Planmeca Viso User s manual...
Страница 110: ......
Страница 111: ......