24 Operating instruments
24.1 Instrument logic
NOTE
If you use similar instruments at the same time, keep them in their own
places. The memory operation does not recognise the parameters and
settings of the similar instruments if their places are changed.
The dental unit contains an instrument logic with the principles described
below. The active instrument can be operated with the foot control and its
settings are displayed on the control panel and they can be changed.
If any settings are modified, the unit saves the settings when the instrument
is returned to the instrument holder.
When the active instrument is operated, other instruments can be away from
the instrument console, but they can not be operated.
The syringe can be operated independently of other instruments at any time.
1. When the dentist picks up instrument 1 from the instrument console, it
becomes the active instrument. The dentist can drive it with the foot
control pedal (a short push to the right/left or down is enough).
2. While the dentist drives instrument 1, the assistant can pick up
instrument 2 for preparation. Picking up instrument 2 does not change
the active instrument.
3. The dentist returns the active instrument (instrument 1) to the instrument
console.
4. The assistant returns instrument 2 to the instrument console.
5. The dentist activates instrument 2 by picking it up from the instrument
console. After this, it can be driven with the foot control as above.
The dental unit stores the settings of each instrument for each instrument
slot separately. When an instrument is inserted into a specific instrument slot
for the first time, the dental unit loads factory default settings for the
instrument. When an instrument is inserted into an instrument slot where it
has been previously, then the dental unit loads the previously saved
instrument settings. The dental unit does not automatically copy instrument
settings from one instrument slot into another.
Example:
The user inserts a Minendo micromotor into the 2nd instrument slot. The
dental unit loads factory default settings for the instrument. The user then
modifies the instrument settings and the dental unit stores the settings for
instrument slot 2 when the user puts Minendo back into the instrument
holder. The user then removes Minendo from instrument slot 2 and inserts a
turbine into the same slot. The dental unit loads factory default settings for
the turbine. The user modifies the settings of the turbine and the dental unit
stores the settings for instrument slot 2. The user then removes the turbine
from the slot and reinserts Minendo. The unit loads the modified and saved
settings for Minendo. The user then inserts the turbine into the 3rd
instrument slot and the unit loads factory default settings for the instrument.
Intelligent four-handed dentistry
In four-handed dentistry it is important to enable a smooth preparation and
exchange of instruments between the dentist and the assistant. Therefore,
24 Operating instruments
136 Planmeca Sovereign Classic
User's manual
Содержание Sovereign Classic
Страница 1: ...PlanmecaSovereign Classic user s manual EN 10032652...
Страница 9: ...35 7 RF Exposure 285 Table of contents User s manual Planmeca Sovereign Classic...
Страница 10: ...Table of contents Planmeca Sovereign Classic User s manual...
Страница 27: ...7 For your safety User s manual Planmeca Sovereign Classic 17...
Страница 290: ...Optimal space requirements 34 2 5 Side view 34 Technical information 280 Planmeca Sovereign Classic User s manual...
Страница 291: ...34 2 6 Top view 34 Technical information User s manual Planmeca Sovereign Classic 281...
Страница 292: ...34 Technical information 282 Planmeca Sovereign Classic User s manual...
Страница 296: ......
Страница 297: ......