Chapter C - HELP & ERROR MESSAGES
Planmeca Proline XC X-ray unit C-9
ERROR MESSAGE EXPLANATIONS
Technical Manual
3.6 Errors 50-57 (connector or cable errors)
Er 50 Temperature sensor shorted
Before each exposure and after pressing READY or kV, mA adjustments the temperature of
the tube head is measured. If the result is too low it is assumed that the sensor is short
circuiting.
Sensor broken. Measure the resistance that should be 1000 ohms at 25 °C. If not near,
replace.
If disconnecting the cable to cassette end limit sensor board cures this error, then this cable
and/or sensor board is partially shorting the + 15V.
Check in service settings mode the measured temperature. If still low with cables discon-
nected, faulty generator processor board.
Er 51 Temperature sensor open (overpressure when applicable)
Before each exposure and after pressing READY or kV, mA adjustments the temperature of
the tube head is measured. If the result is too high it is assumed that the sensor is open cir-
cuit. In some models there is an overpressure sensor in the tube head, connected in series
with the temperature sensor and breaking the circuit at overpressure. Also, if the tube head
and generator processor grounds are not properly connected to the protective earth this error
will be indicated.
Check the earthing cables.
If seems ok by resistance measurement (about 1000 ohms at 25°C) then the generator proc-
essor board is faulty.
Er 52 mA or kV feedback cable open
Before each exposure and after pressing READY the kV and mA feedback signals are meas-
ured. If the result is too high it is assumed that the cables are not properly connected.
If seems ok by resistance measurement (kV about 16 k
Ω
, mA about 383R), measured at the
generator processor board inputs while cables connected, the board is faulty.
Er 57 Exposure button activated
If the exposure circuit is activated when the unit is turned on or service settings mode is
entered, Er 57 occurs. This is to prevent unintentional exposures.
3.7 Errors 60-61 (generator processor and AEC related errors)
Er 60 Supply voltage ± 12 too low
Generator processor board analog circuits have operating voltages of ±12V.
If these are below about 11 V this error is indicated.
If disconnecting the cable to cassette end limit sensor board cures this error, then this cable
and/or sensor board is shorting the + 12 V.
Otherwise faulty generator processor board.
Содержание Proline XC
Страница 1: ...technical manual 10010942_10 EN...
Страница 2: ......
Страница 8: ...TABLE OF CONTENTS 6 Planmeca Proline XC with Dimax3 Technical manual...
Страница 24: ...Chapter A GENERAL TECHNICAL DATA A 16 Planmeca Proline XC X ray unit EMC INFORMATION Technical Manual...
Страница 42: ...Chapter B GRAPHIC USER INTERFACE B 18 Planmeca Proline XC X ray unit SERVICE SETTINGS Technical Manual...
Страница 236: ...Chapter H PARTS REPLACEMENT REPAIR H 14 Planmeca Proline XC X ray unit REPLACING THE SOFTWARE Technical Manual...
Страница 253: ...Planmeca Proline XC X ray unit J 1 Chapter Technical Manual J WIRING DIAGRAM...
Страница 254: ...Chapter J WIRING DIAGRAM J 2 Planmeca Proline XC X ray unit Technical Manual...
Страница 256: ......
Страница 257: ......