Chapter B - GRAPHIC USER INTERFACE
B-12 Planmeca Proline XC X-ray unit
SERVICE SETTINGS
Technical Manual
•
The primary collimator limit sensor operation. When operated by the motor movement the
active time of the sensor is rather short and is not necessarily always visible on this indica-
tor. (Only in Planmeca Proline XC Pan/Ceph X-ray unit).
•
Up/down motor limit sensor
•
Temple support motor limit sensor
•
Layer adjust motor limit sensor
•
Rotating arm end point limit sensor
•
Rotating arm start point limit sensor
•
Cassette holder start (ready) point limit sensor
•
Cassette holder end (return) point limit sensor
2.5 Error history
To make the equipment failure analysis easier an error history display has been provided.
This error memory contains the last 50 errors occurred (not Er 00). The list is shown in chron-
ological order (by date/time). The oldest error messages are dropped out from the list if the
list is full and there are new errors that are stored. Browse the error list with the arrow fields.
NOTE
The error history list can be cleared on the Factory preset window by touching the
Exposure counter button.
Error serial number
Date
Time
Error code number
Содержание Proline XC
Страница 1: ...technical manual 10010942_10 EN...
Страница 2: ......
Страница 8: ...TABLE OF CONTENTS 6 Planmeca Proline XC with Dimax3 Technical manual...
Страница 24: ...Chapter A GENERAL TECHNICAL DATA A 16 Planmeca Proline XC X ray unit EMC INFORMATION Technical Manual...
Страница 42: ...Chapter B GRAPHIC USER INTERFACE B 18 Planmeca Proline XC X ray unit SERVICE SETTINGS Technical Manual...
Страница 236: ...Chapter H PARTS REPLACEMENT REPAIR H 14 Planmeca Proline XC X ray unit REPLACING THE SOFTWARE Technical Manual...
Страница 253: ...Planmeca Proline XC X ray unit J 1 Chapter Technical Manual J WIRING DIAGRAM...
Страница 254: ...Chapter J WIRING DIAGRAM J 2 Planmeca Proline XC X ray unit Technical Manual...
Страница 256: ......
Страница 257: ......